• Only install the optional coolant valve in the
supply line of the exhaust waste vapor con-
denser.
•
Note
: Install the hoses of the cooling system
in a way to avoid the flow / dripping of con
-
densed water onto the pumping unit (espe-
cially cables and electronic parts, see also
IP degree of protection, „Technical data“,
• Ensure that the
coolant outlet tubing
is always unob-
structed and that it cannot get blocked.
• Maximum permissible coolant pressure at the exhaust
waste vapor condenser: 87 psi (6 bar) absolute. Outlet
flow must always be unhindered.
• Comply with the maximum permissible coolant pres
-
sures of additional components in the coolant circuit
(e.g., coolant valve).
• Avoid overpressure in the coolant circuit (e.g., caused
by blocked or squeezed coolant hoses).
✘
✔
EX
IN
Permissible range of coolant temperature at the exhaust
waste vapor condenser: 5°F to 68°F (-15°C to +20°C)
Check hose connections prior to starting operation of the
cooling system.
Check coolant hoses regularly during operation.
NOTICE
MD 1C (AC motor)
Plug in the power cord.
• Check the power source and the pump’s
rating plate to be sure that the power source
and the equipment match in voltage, phase,
and frequency.
Switch the pump on.
Electrical connection
O
➨ I
Summary of Contents for MZ 1C
Page 31: ...page 31 of 92 MZ 1C 3 2 7 10 11 6 9 8 ...
Page 32: ...page 32 of 92 MD 1C AK EK 10 3 6 1 8 2 9 7 15 16 17 13 11 8 14 19 18 ...
Page 33: ...page 33 of 92 MD 1C VARIO SP 7 10 6 12 11 8 Ensure sufficient venting of the pump ...
Page 64: ...page 64 of 92 4 3A 5 MD 1C size 2 2x 1 17 mm 14 mm max 1 4 ...
Page 66: ...page 66 of 92 6 7 8 4 6x Clean ...
Page 67: ...page 67 of 92 9 Replacing the diaphragm 10 11 20696828 20696838 ...
Page 68: ...page 68 of 92 12 1 2 3 13 14 Pay attention to number of washers ...
Page 69: ...page 69 of 92 15 16 17 Pay attention to number of washers ...
Page 70: ...page 70 of 92 18 1 2 19 20 4x Replacing the valves and assembling the pump heads ...
Page 71: ...page 71 of 92 21 22 23 ...
Page 73: ...page 73 of 92 27 28 1 2 28A size 2 2x 17 mm 10 mm MD 1C AK EK 5 1x MD 1C ...
Page 74: ...page 74 of 92 29 30 31 Assembling the connecting hose MD 1C AK EK 2x ...
Page 77: ...page 77 of 92 5 6 4 20638821 20638821 ...
Page 89: ...page 89 of 92 ...
Page 90: ...page 90 of 92 ...
Page 91: ...page 91 of 92 ...