61
20901544_E12C+16C_NT Vsel_V1.2_260922
Cleaning and maintenance
7.3.4
7.3.4 Change the diaphragms and valves
Change the diaphragms and valves
Disassemble the next housing sections
1
2
1.
Unscrew the 4 screws from the
rear housing section; hex key
size 4.
2.
Remove the housing section
and set it aside.
3
4
3.
Unscrew the screws from the
side panel retaining plate; hex
key size 4.
4.
Route the cable out of the
recess.