![vacuubrand MD 12C NT VARIO o.C. Instructions For Use Manual Download Page 56](http://html1.mh-extra.com/html/vacuubrand/md-12c-nt-vario-o-c/md-12c-nt-vario-o-c_instructions-for-use-manual_3201363056.webp)
56
20901544_E12C+16C_NT Vsel_V1.2_260922
Cleaning and maintenance
7.3.2
7.3.2 Preparation
Preparation
Disassemble the controller and base
4 .3 Controller base on page 30
4 .3 Controller base on page 30
Disassemble the device and housing sections
1
2
1.
Switch the diaphragm pump off
and unplug the power plug.
2.
Remove the cap from the gas
ballast.
3
4
3.
Unscrew the 4 screws from the
front housing section; hex key
size 4.
4.
Remove the front housing sec
‑
tion and set it aside.
Disassemble the