33
20901544_E12C+16C_NT Vsel_V1.2_260922
Installation and connection
4.4.2
4.4.2 Outlet connection (EX)
Outlet connection (EX)
WARNING
Risk of bursting due to overpressure inside the
outlet line.
Inadmissibly high pressure in the outlet line can cause
the vacuum pump to burst or damage seals.
>
The outlet line (exhaust gas, gas outlet) must always
be clear and non‑pressurized.
>
Always route the exhaust gas hose with a fall or take
measures to prevent condensate from flowing back
into the vacuum pump.
>
Observe the maximum admissible pressures and
pressure differences.
Connect the exhaust gas hose
1
2
(c) (d)
1.
Take a vacuum hose
(c)
, d
i
15 mm diameter.
2.
Slide the outlet hose onto hose nozzle
(d)
and route the hose
into a fume hood if necessary. If necessary fix the outlet hose,
e. g., with a hose clip.
Example
Exhaust gas connec
‑
tion at the outlet EX