29
20901544_E12C+16C_NT Vsel_V1.2_260922
Installation and connection
Installing the vacuum pump
>
Place the vacuum pump on a stable, non‑vibrating, level sur
‑
face.
5 cm
5 cm
5 cm
>
When installing in lab furniture, maintain a minimum distance
of 5 cm (2 in.) to adjacent objects or surfaces.
>
Prevent heat accumulation and ensure sufficient air circula
‑
tion, especially in closed housings.
Observe limitation of use
Limitation of use
(US)
Ambient temperature
10
–
40 °C
50‑104 °F
Max. altitude
2000 m
above sea level
6562 ft
above sea level
Minimum distance to
adjacent parts
5 cm
2 in
Relative humidity
30‑85 %, non‑condensing
Protection class
IP 40/IK 08
Prevent condensation or contamination from dust, liquids, or corrosive
gases.
>
Note the IP protection class. IP protection is only guaranteed
if the device is appropriately mounted and connected.
>
For connection also note the rating plate data and chapter
8 .1 .1 Technical data on page 78 .
8 .1 .1 Technical data on page 78 .
IMPORTANT!
Example
Sketch of
minimum distances
in lab furniture
Observe limitation of
use
IMPORTANT!