20901544_E12C+16C_NT Vsel_V1.2_260922
Cleaning and maintenance
5
6
5.
Open the hose clip on the
molded hose leading to the
outlet; flat‑head screwdriver
size 1.
6.
Unscrew the 2 screws from the
outlet holder; hex key size 3.
7
7.
Pull the outlet holder, together
with the hose, off the hose
nozzle.
5
Diaphragm pump is pre
‑
pared
Disassemble the
outlet hose