2.2.10 In Case of Trouble
Many difficulties encountered when setting up a new tester are
caused by simple things. Address the items in the checklist below
to ensure the tester is being operated correctly.
Setup Checklist
• Is the power supply connected to the back panel and is it
receiving power (the power supply’s green light is lit)?
• Is supply air connected to Sprint iQ?
• Is the supply air clean and dry? Air that contains moisture,
oil, or hard particulate
will
damage Sprint iQ.
• Is the supply air within the correct pressure range?
• Has the pressure regulator changed from when the test
pressure was set? A change will cause pressure errors.
• Has the LOCKOUT keyswitch been set in the appropriate
position for the desired task (Run vs. Program mode)?
• Is the product to be tested attached to the test port during
all flow tests? Without product, the resulting high flow
could damage the flow sensor.
• Has the calibration been changed? Do
not
enter calibration
mode unless intending to reset calibration data. The
necessary standards and instruments to complete the task
must be at hand.
• Are engineering units set in the correct units of measure?
If the problem persists, call Uson for help.