Calibration BP:
639.4 mmHg
Raw=2753 Offset=0
Pressing
YES
will allow you to save results, pressing
NO
will discard the
results.
YES=Save NO=Cancel
Calibration BP:
639.4 mmHg
Raw=2753 Offset=0
42
The next screen will confirm that the barometric calibration results have
been saved to the memory card. Press
ENTER
to return to the Calibration
Menu.
YES=Save NO=Cancel
6.3 Flow Calibration for Module C
Pressing
←
F3
at the Calibration Menu will allow you to step-through a
three-point flow calibration. Refer to Section 4.12 for a complete flow
calibration software menu map.
Calibration Mod:[1]
NO=Back ENTER=Next
Select
Mod: 1
for doing a flow calibration on the Module C, and
ENTER
to
begin.
WARNING
A warning will be shown reminding you that a leak check should be
performed before an audit
A leak check should
always precede a
calibration
WARNING
Continue with
calibration?
The next screen warns you not to proceed with calibration unless a leak
check has been performed. Pressing
YES
will continue.
NO=Back YES=Continue
Cal pt: 1 Mod:[1]
Flow set pt: 19.80
The next screen will show the Calibration Point (1 of 3), the Module (1),
and the Flow set point of 19.80. Press
ENTER
to continue the process.
ENTER=Next
Cal pt: 1 Mod:[1]
Connect Reference
Flowmeter Now!
The next screen will prompt you to install the reference flow meter.
NO=Cancel ENTER=Next
Cal pt: 1 Mod:[1]
<MFC Warmup>
The mass flow controller will warm up for a few seconds
Cal pt: 1 Mod:[1]
The pump will warm up for 300 seconds (5 minutes). Pressing
ENTER
at
any time during the pump warmup will stop the pump warmup and advance
to the next screen.
<Pump Warmup 300>
Next, the Gain, Offset, Raw, and Flow values are shown for the selected
module. Again, pressing
ENTER
will advance to the next screen.
Cal pt: 1 Mod:[1]
Gain=6.00 Off= 0.00
Raw=2800 Flow= 19.77
NO=Cancel ENTER=Next
Again, pressing
ENTER
will advance to the next screen.