4.6.4.2 Flow Audit
WARNING
31
Pressing
↓
F2
at the Audit Menu will bring you to the Flow Check
procedure. Initially, a warning will be shown reminding you that a leak
check should be performed before an audit.
A leak check should
always precede an
audit.
4.6.4.3 Temperature Audit
Audit Temperature
Temperature(C)= 25.4
F1:+/- F2:C/F
Pressing
←
F3
on the Audit Menu brings you to the Temperature Audit
screen
Ref. Temp(C):?
4.6.4.4 BP Audit
Audit BP
Pressing
→
F4
on the Audit Menu brings you to the BP Audit screen.
BP (mmHg)= 643.9
Ref. BP(mmHg):?
4.7 Main Menu (3 of 5)
4.7.1 Overview
F1=View protocol
F2=View ID codes
F3=View operators
The third Main Menu allows for you to view sampling protocol, sampler ID
codes, or view the initials of the site operator.
F4=More ENTER=Back
4.7.2 View
Protocol
Sampling Protocol
1 in 3
Sample Time/Hour 60
Pressing
↑
F1
allows you to view the sampling protocol, as shown. In this
example, the Sample Days are shown as 1 in 3, and the Sample
Time/Hour is shown as 60.
ENTER=Back
4.7.3 View
ID
codes
LOC:0000000000
Q:Q0000000
Comp:I000000I
SN:0000 ENTER=Back
Pressing
↓
F2
at the third Main Menu enables you to once again view the
ID codes that were shown on startup. Here, the display shows the Location
Code (LOC), the 15-character Chain of Custody (Q), the 15-character
Module ID (Comp), and the Serial Number (SN). The
ENTER
key will
return you to the Main Menu.