Programming and calibrating
48
To do so, carry out the following steps:
1. Hold down the
key for approx. 2 seconds:
2. Actuate the pump nozzle over a collecting tray until the feed pump stops
automatically. Hold the pump nozzle
at the same height as when dosing
later on
to achieve the greatest possible precision.
3. After the dosing duration of 5 seconds, the display
shows the view opposite:
The left value shows the measured value of the current measurement, the right
value is the calibration value used for calculation.
4. Measure the conveyed feed amount.
5. Press the
key approx. 2 seconds.
6. Use
/
to enter the measured feed amount
(e.g. 3.3 litres)
7. Briefly press
to exit the calibrate
settings:
8. Briefly press
or
to exit the calibrate menu.
7.2.3
Temperature calibration
The temperature regulation of the Milk Shuttle is calibrated precisely before the
device leaves the factory. Recalibration is only then necessary if the true
temperature of the feed deviates from the displayed actual temperature.
To do so, carry out the following steps:
1. Hold down the
key for approx. 2 seconds.
2. Briefly press the
key:
3. Use a thermometer to measure the deviation of the feed temperature from the
target temperature.
Information
You can measure the feed amount alternatively in kg or in litres. The chosen unit
(kg or litre) must then be used when entering the required volumetric delivery.
1:Pump
03.200
168
1:Pump
03.200
167
168
INFO
SET
+
-
SET Pump
>03.200<
167
168
SET
SET Pump
>03.300<
167
168
ESC
INFO
2
2:Temperature
14.5 C
+0.0
Summary of Contents for MilkShuttle
Page 6: ...About these operating instructions 6 ...
Page 28: ...Start up 28 ...
Page 41: ...Operation 41 Figure 6 7 Moving on slopes 9 x ...
Page 42: ...Operation 42 ...
Page 56: ...Programming and calibrating 56 ...
Page 78: ...Maintenance Repairs 78 ...
Page 80: ...Spare parts 80 ...
Page 84: ...Mixing tables 84 ...
Page 86: ...Index 86 ...
Page 87: ...87 Notes ...