CALIBRATION
29
C
A
LI
B
R
A
TI
O
N
Chapter
4
(calibration error)
The initial dead load exceeds the zero adjustment range.
Connect a resistor of an appropriate value between the +EXC and -SIG terminals of the load cell,
shift the zero point, and then perform zero calibration again.
(calibration error)
The initial dead load is negative.
Check to see if the load cell is put under load in the correct direction and check the wiring of +SIG
and -SIG for reversal, and then perform zero calibration again.
For reference, relationships between connected resistance values and input
signals are listed below.
- The numerical values in this table assume that one 350-ohm load cell is used.
In the case where four load cells are connected in parallel, the sensitivity
decreases to 1/4; therefore, the resistance values should be considered as 1/4.
- The temperature coefficient of the resistance connected here directly affects
the accuracy of the indicator.
Use resistance of 50ppm/
℃
or higher (approx. 5 - 10ppm/
℃
recommended) at
least.
Resistance value
Input conversion strain
Calculated value
Approximate value
μ-STRAIN
mV/V
875
kΩ
866
kΩ
200
0.1
437
kΩ
442
kΩ
400
0.2
291
kΩ
294
kΩ
600
0.3
219
kΩ
221
kΩ
800
0.4
175
kΩ
174
kΩ
1000
0.5
146
kΩ
147
kΩ
1200
0.6
125
kΩ
124
kΩ
1400
0.7
109
kΩ
110
kΩ
1600
0.8
97
kΩ
97.6
kΩ
1800
0.9
87.3
kΩ
86.6
kΩ
2000
1.0
79.4
kΩ
78.7
kΩ
2200
1.1
72.7
kΩ
73.2
kΩ
2400
1.2
67.1
kΩ
66.5
kΩ
2600
1.3
62.3
kΩ
61.9
kΩ
2800
1.4
58.2
kΩ
57.6
kΩ
3000
1.5
54.5
kΩ
54.9
kΩ
3200
1.6
51.3
kΩ
51.1
kΩ
3400
1.7
48.4
kΩ
48.7
kΩ
3600
1.8
45.9
kΩ
46.4
kΩ
3800
1.9
43.6
kΩ
43.2
kΩ
4000
2.0
41.5
kΩ
41.2
kΩ
4200
2.1
39.6
kΩ
39.2
kΩ
4400
2.2
37.9
kΩ
38.3
kΩ
4600
2.3
36.3
kΩ
36.5
kΩ
4800
2.4
34.8
kΩ
34.8
kΩ
5000
2.5
Point