background image

23 of 33 

to shift the cursor to the next position. When 
none of the digit is blinking, key K2 is used 
to execute the password. 

CALIBRAT 

MODE

 

After entering the correct password (0001), 
LCD/OLED shows this prompt. It means now it is 
in calibration mode.  Use K2 to proceed 
forward. 

 
 
 
 

 
 
 
 

APPLY  

ZERO- Gas

 

 

The unit enters in 

“Zero

-

Cal” & display

 prompts 

to apply Zero-Gas or clean air free from the 
target gas. LCD/OLED indicates 30 second down 
count, this is the fixed time allotted for base 
line or zero cal setting. After successful base 
line setting, the LCD/OLED 

will give “Ze

ro-Cal 

Done” prompt. In case if the base line of the 

sensor is not proper the Zero- Cal will fail & 
unit 

will show “Zero

-

Cal fail”

 

 
 
 
 
 

 
 
 

ENTER SPAN 

GAS VALUE 

XXX

 

 

 
 

 

After “Zero cal Done” 

use K2 to ente

r in “Span

-

Cal” & 

prompt will appear for the Span gas (or 

calibration gas) value to enter (which is 
available for the calibration of the unit).  
 
On this prompt, by using key K2, span gas value 
can be incremented / decremented. Key K1 is 
used to toggle between increment/decrement 
mode, this is indicated by blinking cursor at 
the locations.  If correct span gas value is 
entered on pressing K1 further, the blinking 
cursor disappears.  Press K2 to accept the 
entered value and proceed further and select 
span time.  
 

APPLY  

SPAN- Gas

 

 

After entering the span gas value and time, the 
unit prompts for applying the span gas. 
LCD/OLED also shows the down count. After 
successful span setting, LCD/OLED gives 

“Span

-

Cal Done” prompt. In case if the span setting 

of the sensor is not proper the Span- Cal fails 
& unit 

will show “Span

-

Cal failed”. If span cal 

fails then the previous calibration data of the 
sensor will be restored back and fresh attempt 
of the sensor calibration need to be done. 
On successful calibration, the calibration date 
automatically gets updated. 

Summary of Contents for 500DT

Page 1: ...1 of 33 UNIPHOS 500DT Digital Transmitter Uniphos Envirotronic Pvt Ltd Readymoney Terrace 167 Dr Annie Besant Road Worli Mumbai 400 018 India INSTRUCTION MANUAL...

Page 2: ...y Signal 4 20mA 13 Settings 15 C Operation 19 Gas alarm Conditions Advice 20 Fault alarm Conditions Advice 21 D Calibration 22 Guidelines 22 Calibration procedure 22 Calibration Diagnostics Protection...

Page 3: ...ly ensuring that the components are not damaged otherwise the approval may be adversely affected wrong readings could occur and persons relying on this product for their safety could sustain serious p...

Page 4: ...gnal which can be connected to any control system like DCS PLC etc Designed with non volatile memory the transmitters have total memory retention capacity There is provision to have mute function of t...

Page 5: ...es easy installation and the fastest routine operation by removing the need for hot work permits in hazardous areas Using easy to replace plug in sensor modules downtime is also reduced and on going c...

Page 6: ...6 of 33 B Installation Various parts of the gas transmitter are shown in the following diagram Figure 1 500 DT Transmitter overview...

Page 7: ...we suggest placing the sensor approximately 12 30 cm above the level of a potential leak source or near the roof in indoor installation For gases heavier than air we suggest placing the sensor 12 30 c...

Page 8: ...r supply signal cable connections Opening on the Left hand side is equipped with 1 2 NPT M20 conduit fitting outlet Use certified cable gland for cable entry It is to be used for the Relay cable entri...

Page 9: ...9 of 33 Figure 2 Outline drawing Front side views with outline dimensions...

Page 10: ...e a resistance of 100 ohm maximum Maximum cable lengths between the controller transmitter The table give below is for guidance only Users should calculate maximum distances using actual data for the...

Page 11: ...11 of 33 Wiring Schematic diagram UNIPHOS transmitter can be wired in current source mode as shown in the below wiring schematic diagram Figure 3...

Page 12: ...follow Ref Figure 4 Relay Cable Guideline Optional The transmitter is equipped with one potential free Relay optional contacts Relay connections are labeled NO normally open NC normally closed C commo...

Page 13: ...e as shown in the figure 4 Connect 3 wires as following Positive Supply 24V 24 VDC to the labeled terminal Signal 4 20mA 4 20mA loop signal to the labeled terminal Negative Supply AGND Ground wire to...

Page 14: ...14 of 33 Figure 4 System wiring diagram Note The Armored cable Screen is to be terminated on to the Earthing point provided on interior of the Enclosure...

Page 15: ...activation Alm1 threshold Level Value This is Low alarm activation level Alm2 Short term Threshold Level Value This is High alarm activation level Mask This is factory setting Please do not change it...

Page 16: ...t towards left position After the last digit blinking cursor disappears press K2 to accept changes and go to the next prompt 3 ALM2 xxxx In a similar fashion the set ALM2 and move to the next prompt 4...

Page 17: ...ovision is given to activate deactivate the Relays operations site when required Oxygen Selection This provision is given to enable disable the Relays Alarm logic for Oxygen gas To go into the paramet...

Page 18: ...Date Key K2 is used to increment the digit where the cursor is positioned and key K1 is used to shift the cursor to the next towards left position After the last digit blinking cursor disappears pres...

Page 19: ...ion 1 UNIPHOS 500 DT Model number of the product 2 KwkSense Lite Name of the product 3 VERSION XX XX XXa Current software version 4 ERROR NO SENSOR Error message if sensor is not connected 5 CAL DATE...

Page 20: ...larm relay triggers the connected device Visual Indicator LED turns to blue color Numeric LCD OLED display indicates the gas concentration currently present in ppm LEL or V V HIGH GAS ALARM The gas co...

Page 21: ...t conditions Fault condition 4 20mA Output LED Visual Indicator LCD OLED display Sensor self test fail At start up sensor self test failed 2 mA Visual Indicator LED turns to red color Numeric LCD OLED...

Page 22: ...chemical sensors Tygon may poison a catalytic combustible sensor over time Higher flow rate of calibration gas could give incorrect reading due to pressure build up over the sensor Calibrate the senso...

Page 23: ...pan gas or calibration gas value to enter which is available for the calibration of the unit On this prompt by using key K2 span gas value can be incremented decremented Key K1 is used to toggle betwe...

Page 24: ...alibration Wait for the transmitter to return to normal operation Replace the empty gas cylinder with a full cylinder repeat process Calibration gas concentration is too low or too high Calibration ga...

Page 25: ...to 2 months or a long interval Other Settings Take the magnetic wand to K2 system will prompt to enter Password Enter 0004 and proceed forward for loop current settings 4 20mA Please see the below tab...

Page 26: ...g the Sensor The sensor is secured in SS adaptor with sinter or filter screen depending on gas Clean the sinter filter only with a dry brush being careful not to clog the screen Clearing the Sensor Th...

Page 27: ...LCD OLED showing Error No sensor sensor is not properly plugged in the sensor housing Remove the sensor re plug it again carefully Transmitter LCD OLED showing Error Sensor drift Or Over Range Sensor...

Page 28: ...n the sensor module a Battery back up is not necessary to retain values in the event of Power outages Loop Resistance 600 ohms maximum for 4 20 mA loop current signal Self Test Automatic self test of...

Page 29: ...ument 1 year Sensor 1 year on proportional basis Approval Optional 1 ATEX Ex db IIC T4 Gb Tamb 20 C to 85 C 2 IECEx Ex db IIC T6 Gb Tamb 20 C to 40 C Ex db IIC T5 Gb Tamb 20 C to 55 C Ex db IIC T4 Gb...

Page 30: ...0 PPM 0 1 PPM 040A SO2 sensor electro chemical Sulphur Dioxide 0 20 PPM 0 1 PPM 050A Phosphine sensor electro chemical Phosphine 0 10 PPM 0 01 PPM 055A Phosgene Sensor electro chemical Phosgene 0 1 PP...

Page 31: ...9C Flammable sensor Pellistor Acetylene 0 100 LEL 1 LEL 099D Flammable sensor Pellistor Butanol 0 100 LEL 1 LEL 099E Flammable sensor NDIR Methane 0 100 LEL 1 LEL 099F Flammable sensor NDIR Methane 0...

Page 32: ...32 of 33 Notes...

Page 33: ...9909994042 91 7574839945 Email Id info uniphos envirotronic com After Sales Support email id service uniphos envirotronic com Website www uniphos envirotronic com Issue 0 Rev 07L Date 08 02 2022 Prepa...

Reviews: