79
7. Performance Parameters
This chapter lists spectrum analyzer’s technical parameters and general technical specifications.
Unless otherwise specified, technical parameters are applicable to following conditions:
z
Instrument has been warmed up for 30 minutes before use.
z
Instrument is within calibration period and self-calibration has been executed.
z
Typical value of tolerance is not provided. Typical value means parameters not guaranteed by product.
Frequency
Frequency range
Frequency resolution
Frequency reading
accuracy
± (frequency marker reading×frequency benchmark
a1%×span+10%×RBW+0.5×[span/(sweep point-1)]+1Hz)
Internal benchmark
(10MHz)
Aging rate <1ppm/y
Temperature drift <0.5ppm (15
ć
to 35
ć
)
Resolution bandwidth (RBW)
Range (<10Hz as option)
1Hz to500kHz (continuous stepping from 1 to 10), 1MHz, 3MHz
Selectivity (60dB/3dB)
RBW
500kHz <5:1 typical value (digital realized, close to Gaussian shape)
Accuracy
Typical value
Video bandwidth (VBW)
10Hz to 3MHz