– PROBES INSTALLATION

 

– (General Procedures). 

 

 Unless specified and instructed otherwise in the order, refer to the probe data sheet  

of the model used.  

 

 A dual-chord spool must always be constructed according to our guide drawing. Once 

the structure has been made, verify the sizes “D 

int

, L

1

 & L

2

 , D

1

 & D

2

 , d

1

 & d

2

” as 

built 

Enter the results onto a sheet which will be used during the parameter adjustment. 
! The final accuracy of the measurement depends on it. 

 

  The  installation  of  the  probes  in  a  conduit  should  allow  for  the  best  the  straight 

lengths,  particularly  upstream,  but  also  downstream,  in  order  to  benefit  from  a 
predictable and stabilized hydraulic profile. Provide upstream a minimum of 15 times 
the pipe diameter. 

 
 
 
 
 
 

  Similarly,  to  avoid  the  nuisance of  deposits  or  accumulation  of gas,  planes  close  to 

vertical (+/- 30°) will be avoided. A horizontal plane is a good choice excepted when 
the conduit is vertical. 

 
 
 
 
 
 

  For  intrusive  probes,  adjust  the  insertion  then  the  alignment  or  orientation  of  the 

probes for the same pair (> parallel faces). 

 
 
 
 
 

  For  external  probes,  the  freedom  offered  should  not  prompt  the  user  to  ignore                

the above rules. 
 However,  the  choice  of  direct  (\),  reflex  (V)  or  even  (N  or W)  installation  mode  will 
depend on the decision regarding the conduit (state) and the fluids to be measured.  
 Therefore,  we  advise  initially  a  provisional  installation  known  as  a  validation 
installation (coupled with gel or grease) and final assembly (supports, solid coupling) 
once the conditions have been validated. 
 

Since the MiniSonic calculates for you the distance required between the probes, the 

setting up operations will, in theory, take place once the converter parameters have 
been set 

 Particular attention will be paid to cleaning the conduit at the probe location. 

 
 
 

D.S. 

 

    D 

Summary of Contents for Minisonic_600-2

Page 1: ...ual chord Minisonic_600 2 Minisonic 2000 dual chord Minisonic_2000 2 B timent TEXAS ragny Parc 9 All e Rosa Luxemburg 95610 RAGNY FRANCE T l 33 0 1 30 27 27 30 Fax 33 0 1 30 39 84 34 www ultraflux net...

Page 2: ...0 B and 2000 B for dual pipe flow measurement configurations Speed 1 or 2 is for open channel flow velocity measurements This manual is specifically concerned with an element from the family and has b...

Page 3: ...ersion Ex d ATEX Version 4 Wiring diagram 5 Converter Assembling Instructions Mechanical installation Electrical wiring 6 Probes installation 7 Setting and commissioning 8 Detailed Menu layout Measure...

Page 4: ...with tapping at an oblique angle and straight ended probes Or with taps perpendicular to the pipe but with bevel end probes Good geometrical control is essential The construction of a measuring spool...

Page 5: ...afety risks N A Only low voltage 60 V Concerning EMC EN 61000 EN 55022 et EN 50204 Concerning tightness IP EN 60529 Concerning ATEX Directive 94 4 CE EN 60079 0 General rules EN 60079 1 Ex d enclosure...

Page 6: ...C T6 IP 67 See also the specific notice NT 219 B Each touch of the keypad has the same function as above The two glands near the front are reserved for probe cables These connections request the use o...

Page 7: ...8 7 6 5 4 3 14 13 12 BR5 2 1 1 100 mm 4 WIRING DIAGRAM FOR MINISONIC 600 2 2000 2 Low voltage supply See also NT 218 if the GP 01 90 230 V AC 24 V DC if installed The internal cables must be kept as s...

Page 8: ...he limits V 100 V I 0 1 A R ON 10 Ohm Each relay Re 1 terminals 7 8 or Re 2 terminals 9 10 will transmit the chosen status information in the General Parameters menu 4 20 mA analog outputs BR 2B termi...

Page 9: ...mulation of gas planes close to vertical 30 will be avoided A horizontal plane is a good choice excepted when the conduit is vertical For intrusive probes adjust the insertion then the alignment or or...

Page 10: ...ettings using the PC software Refer to the manual supplied with the software CD ROM The fields of the main values to be entered are listed below under Heading 8 The software offers extended configurat...

Page 11: ...els 1 2 and also averaged values Physical measurements time of flight T and difference dT channels 1 2 Totalizers Reset possible by simultaneously pressing and Date and time Change by PC Software Hydr...

Page 12: ...t action to take place An operation will be requested allowing a provisional code to be calculated A second press on will provide access to the existing Languages choice field 8 3 Pipe Settings menu A...

Page 13: ...f Coefficient KH or following data to do its calculation If MANUAL entry of coefficient depending on profile Laminar 1 333 or Turbulent 1 02 to1 08 If AUTO entry of parameters Fluid viscosity Pipe rou...

Page 14: ...Otherwise If MANUAL Gain adjustment If AUTO Margin Over amplification compared to critical gain Go If SM choice position of probes depending on geometrical measurements as distance between probes fac...

Page 15: ...gnment If copy of TOT 1 or TOT 2 relay closing time F maxi If Threshold enter value Q Reset to Zero of TOT t1 t2 This function may be prohibited Parameter settings of two outputs 4 20 mA SA 1 SA 2 by...

Page 16: ...actory Settings menu normally reserved to Ultraflux It permits to adjust a coefficient which optimises the compensation with Clamp on probes value 0 6 to1 6 Another parameter named Delta T band permit...

Page 17: ...ment displayed and transmitted meets your expectations Case 2 The measurement operates but may lack accuracy or is unstable Case 3 The measurement is not reliable or the MiniSonic is displaying a Faul...

Page 18: ...Diameter Flow Rate Assembly and Choice of probes is too low to claim any measurement accuracy Is it possible to revise these choices The measurement conditions reveal poor control of transitions from...

Page 19: ...T 4 Echo Alternating signal peak to peak 4 Volt max 0 65 V negative threshold voltage PT 0 0 Volt Ground PT 2 Synchro Trig on Positive edge 0 to 5 Volt PT 1 Selection windows Echo must be inside PT 3...

Page 20: ...NT 208D GB1 20 20 11 APPENDIX Calibration file example with a MiniSonic 2000 2 on a flow spool with wetted probes SM...

Reviews: