NT 208D GB1
19 / 20
If the problems are encountered only on one channel, it is easy to swap both
channels and see if the issue is with the application or the MiniSonic.
10 - INVESTIGATIONS:
If there is a difficulty or a problem, logical analysis can be assisted by the
implementation of expert apparatus:
Either an oscilloscope: the MiniSonic has test points for the purpose:
From low to high referring to wiring diagram § 4:
PT 4 = Echo
– Alternating signal, peak to peak 4 Volt max., (-0.65 V negative
threshold voltage).
PT 0 = 0 Volt
– Ground
PT 2 = Synchro.
–Trig on Positive edge (0 to 5 Volt).
PT 1 = Selection windows .Echo must be inside.
PT 3 = Emission < > Reception transit time.
PT 0 = 0 Volt.
Or a MiniSonic P or a DigiSonic E / P or an UF 810-P portable flowmeter.
Or measuring instruments for dimensional or alignment measurements
SPARE PARTS
:
Ultraflux offers possibilities of purchasing spare cards or, as long as acceptance is
granted, the standard replacement of equipment in the event of failure.
Contact the Ultraflux commercial services or the regional dealer.