background image

 

NT 208D GB1

 

 

 

12 / 20

 

8-2

 

: « Calibration Allowed »

 

Menu : 

 

 Access = press   

F

 once from the Measurement Display menu 

 

 If the display is:  

Enter this code (1 to 65535) 

 

 If the display does not request 

“Code?” there is none (code = 0).  

 

 Press    

   

 Enter a code: 

 
 

! This code will become active after de-energizing.  

 

Remember  to  make  a  note  of  it.    It  will  be  required  for  any  subsequent 
intervention. 

 
 

 In case of loss, contact Ultraflux giving the following information: the serial number of 

your MiniSonic MK4*/xx/xx/xxxx as it appears during energizing and the date or dates 
on  which  you  want  action  to  take  place.    An  operation  will  be  requested  allowing  a 
provisional code to be calculated. 

 

 A second press on   

   will provide access to the existing Languages choice field. 

 
 

8-3 : « Pipe Settings » menu : 

 

 

 Access = Press twice on   

F

   or    

F   

once, then    

+

   , and  Enter by   

    

(Operations are common to all the menus). 

 

 

Outside diameter (mm)  
or pipe circumference . 

 
 

 

Chosen pipe material  

  from programmed list :  

 

 If Other: Characterization of this material by its velocity or CM 

 

This  selection  will  resolve  new  or  special  cases  and  also  optimize  the 
parameters  for  composite  pipes  having  several  materials  (steel  or  cast  iron  + 
cement) . 

 

 

Total pipe thickness: 

 
 

 

If Other: 

 
 

ext D = xxxx.x 
ext C = xxxx.x

 

Pipe = 
STEEL , PVC ,OTHER ...

 

Thickness = xx.x mm 

CM = xxxx m/s 

Calibration Allowed : 0 
Code ?

 

 

Access 

 

Code = xxxx 

Summary of Contents for Minisonic_600-2

Page 1: ...ual chord Minisonic_600 2 Minisonic 2000 dual chord Minisonic_2000 2 B timent TEXAS ragny Parc 9 All e Rosa Luxemburg 95610 RAGNY FRANCE T l 33 0 1 30 27 27 30 Fax 33 0 1 30 39 84 34 www ultraflux net...

Page 2: ...0 B and 2000 B for dual pipe flow measurement configurations Speed 1 or 2 is for open channel flow velocity measurements This manual is specifically concerned with an element from the family and has b...

Page 3: ...ersion Ex d ATEX Version 4 Wiring diagram 5 Converter Assembling Instructions Mechanical installation Electrical wiring 6 Probes installation 7 Setting and commissioning 8 Detailed Menu layout Measure...

Page 4: ...with tapping at an oblique angle and straight ended probes Or with taps perpendicular to the pipe but with bevel end probes Good geometrical control is essential The construction of a measuring spool...

Page 5: ...afety risks N A Only low voltage 60 V Concerning EMC EN 61000 EN 55022 et EN 50204 Concerning tightness IP EN 60529 Concerning ATEX Directive 94 4 CE EN 60079 0 General rules EN 60079 1 Ex d enclosure...

Page 6: ...C T6 IP 67 See also the specific notice NT 219 B Each touch of the keypad has the same function as above The two glands near the front are reserved for probe cables These connections request the use o...

Page 7: ...8 7 6 5 4 3 14 13 12 BR5 2 1 1 100 mm 4 WIRING DIAGRAM FOR MINISONIC 600 2 2000 2 Low voltage supply See also NT 218 if the GP 01 90 230 V AC 24 V DC if installed The internal cables must be kept as s...

Page 8: ...he limits V 100 V I 0 1 A R ON 10 Ohm Each relay Re 1 terminals 7 8 or Re 2 terminals 9 10 will transmit the chosen status information in the General Parameters menu 4 20 mA analog outputs BR 2B termi...

Page 9: ...mulation of gas planes close to vertical 30 will be avoided A horizontal plane is a good choice excepted when the conduit is vertical For intrusive probes adjust the insertion then the alignment or or...

Page 10: ...ettings using the PC software Refer to the manual supplied with the software CD ROM The fields of the main values to be entered are listed below under Heading 8 The software offers extended configurat...

Page 11: ...els 1 2 and also averaged values Physical measurements time of flight T and difference dT channels 1 2 Totalizers Reset possible by simultaneously pressing and Date and time Change by PC Software Hydr...

Page 12: ...t action to take place An operation will be requested allowing a provisional code to be calculated A second press on will provide access to the existing Languages choice field 8 3 Pipe Settings menu A...

Page 13: ...f Coefficient KH or following data to do its calculation If MANUAL entry of coefficient depending on profile Laminar 1 333 or Turbulent 1 02 to1 08 If AUTO entry of parameters Fluid viscosity Pipe rou...

Page 14: ...Otherwise If MANUAL Gain adjustment If AUTO Margin Over amplification compared to critical gain Go If SM choice position of probes depending on geometrical measurements as distance between probes fac...

Page 15: ...gnment If copy of TOT 1 or TOT 2 relay closing time F maxi If Threshold enter value Q Reset to Zero of TOT t1 t2 This function may be prohibited Parameter settings of two outputs 4 20 mA SA 1 SA 2 by...

Page 16: ...actory Settings menu normally reserved to Ultraflux It permits to adjust a coefficient which optimises the compensation with Clamp on probes value 0 6 to1 6 Another parameter named Delta T band permit...

Page 17: ...ment displayed and transmitted meets your expectations Case 2 The measurement operates but may lack accuracy or is unstable Case 3 The measurement is not reliable or the MiniSonic is displaying a Faul...

Page 18: ...Diameter Flow Rate Assembly and Choice of probes is too low to claim any measurement accuracy Is it possible to revise these choices The measurement conditions reveal poor control of transitions from...

Page 19: ...T 4 Echo Alternating signal peak to peak 4 Volt max 0 65 V negative threshold voltage PT 0 0 Volt Ground PT 2 Synchro Trig on Positive edge 0 to 5 Volt PT 1 Selection windows Echo must be inside PT 3...

Page 20: ...NT 208D GB1 20 20 11 APPENDIX Calibration file example with a MiniSonic 2000 2 on a flow spool with wetted probes SM...

Reviews: