background image

 

NT 208D GB1

 

 

 

18 / 20

 

 

Case 2

 : In addition to the points mentioned above, the possible causes are: 

 

  Poor control of the installed dimensions, or of the product characterisation or of the 

Zero point : 

  Correct and reanalyse 

 

  An installation constructed with poor hydraulic conditions : 

  If possible, change the location of the probes or spool piece if possible. 

  Otherwise,  for  external  probes,  choose  KHydro  =  Manual  and  adjust  this 

coefficient to obtain the expected flow rate reading. 

  Or, for internal probes, adjust the coefficients CV 1 & 2. 

 

  The range of speed of sound to be measured is too wide for the chosen setup : 

  With S.E. probes, a direct mode ( / ) may be preferable to a reflex mode ( V ) 

  Have you correctly set Co and Delta C range? 

 

 

An unstable fluid causing the ESC mode to restart too often: 

  try the AUTO mode with a reduced margin (

 12 dB) 

 

  The expected measurement requires greater precision : 

The MiniSonic allows the input of a “Linearization” correction law. 

 Plot the error curve and use the software to enter the complementary law (Factory 

Adjustments Menu).  

 

  The  Delta  t  offered  by  the  measurement  conditions  (Diameter,  Flow  Rate, 

Assembly, and Choice of probes) is too low to claim any measurement accuracy. 

  Is it possible to revise these choices?

 

 

  The  measurement  conditions  reveal  poor  control  of  transitions  from  turbulent  to 

laminar configurations: 

  Revise the declared Reynolds Critical and range (LBR) values. 

 

 

Case 3

 : The MiniSonic does not operate: 

 

  There is no display or measurement transmission ( 4 

– 20 mA ) .  

 Is the MiniSonic powered up?  Check with a voltmeter.  

  Has it failed?  If yes, the only solution is to call in our after-sales service. 

 

 

  If the MiniSonic displays INIT constantly, attempt a power cutoff, then re-energize. 

  If the fault persists, contact Ultraflux (Problem with the microprocessor). 

    If  the  fault  disappears,  it  may  come  back  and  could  be  due  to  excessively  high 

impedance in the supply source.      

 

 

If the MiniSonic displays “Flow Rate Fault”, this message does not mean that the 
MiniSonic has failed, but that the ultrasound signals have not reached the expected 
level or are outside the window [Co +/- Delta C]. There are several possibilities : 

  Pipe is empty? 

  Physical problem from site: Two-phase, highly charged or too viscous fluid? 

  Is pipe too old, which does not transmit ultrasonic signals? 

  Unsuitable installation mode = try Direct mode (\) or other sensors. 

  

Incorrect sensor installation (orientation, coupling … ) . 

Summary of Contents for Minisonic_600-2

Page 1: ...ual chord Minisonic_600 2 Minisonic 2000 dual chord Minisonic_2000 2 B timent TEXAS ragny Parc 9 All e Rosa Luxemburg 95610 RAGNY FRANCE T l 33 0 1 30 27 27 30 Fax 33 0 1 30 39 84 34 www ultraflux net...

Page 2: ...0 B and 2000 B for dual pipe flow measurement configurations Speed 1 or 2 is for open channel flow velocity measurements This manual is specifically concerned with an element from the family and has b...

Page 3: ...ersion Ex d ATEX Version 4 Wiring diagram 5 Converter Assembling Instructions Mechanical installation Electrical wiring 6 Probes installation 7 Setting and commissioning 8 Detailed Menu layout Measure...

Page 4: ...with tapping at an oblique angle and straight ended probes Or with taps perpendicular to the pipe but with bevel end probes Good geometrical control is essential The construction of a measuring spool...

Page 5: ...afety risks N A Only low voltage 60 V Concerning EMC EN 61000 EN 55022 et EN 50204 Concerning tightness IP EN 60529 Concerning ATEX Directive 94 4 CE EN 60079 0 General rules EN 60079 1 Ex d enclosure...

Page 6: ...C T6 IP 67 See also the specific notice NT 219 B Each touch of the keypad has the same function as above The two glands near the front are reserved for probe cables These connections request the use o...

Page 7: ...8 7 6 5 4 3 14 13 12 BR5 2 1 1 100 mm 4 WIRING DIAGRAM FOR MINISONIC 600 2 2000 2 Low voltage supply See also NT 218 if the GP 01 90 230 V AC 24 V DC if installed The internal cables must be kept as s...

Page 8: ...he limits V 100 V I 0 1 A R ON 10 Ohm Each relay Re 1 terminals 7 8 or Re 2 terminals 9 10 will transmit the chosen status information in the General Parameters menu 4 20 mA analog outputs BR 2B termi...

Page 9: ...mulation of gas planes close to vertical 30 will be avoided A horizontal plane is a good choice excepted when the conduit is vertical For intrusive probes adjust the insertion then the alignment or or...

Page 10: ...ettings using the PC software Refer to the manual supplied with the software CD ROM The fields of the main values to be entered are listed below under Heading 8 The software offers extended configurat...

Page 11: ...els 1 2 and also averaged values Physical measurements time of flight T and difference dT channels 1 2 Totalizers Reset possible by simultaneously pressing and Date and time Change by PC Software Hydr...

Page 12: ...t action to take place An operation will be requested allowing a provisional code to be calculated A second press on will provide access to the existing Languages choice field 8 3 Pipe Settings menu A...

Page 13: ...f Coefficient KH or following data to do its calculation If MANUAL entry of coefficient depending on profile Laminar 1 333 or Turbulent 1 02 to1 08 If AUTO entry of parameters Fluid viscosity Pipe rou...

Page 14: ...Otherwise If MANUAL Gain adjustment If AUTO Margin Over amplification compared to critical gain Go If SM choice position of probes depending on geometrical measurements as distance between probes fac...

Page 15: ...gnment If copy of TOT 1 or TOT 2 relay closing time F maxi If Threshold enter value Q Reset to Zero of TOT t1 t2 This function may be prohibited Parameter settings of two outputs 4 20 mA SA 1 SA 2 by...

Page 16: ...actory Settings menu normally reserved to Ultraflux It permits to adjust a coefficient which optimises the compensation with Clamp on probes value 0 6 to1 6 Another parameter named Delta T band permit...

Page 17: ...ment displayed and transmitted meets your expectations Case 2 The measurement operates but may lack accuracy or is unstable Case 3 The measurement is not reliable or the MiniSonic is displaying a Faul...

Page 18: ...Diameter Flow Rate Assembly and Choice of probes is too low to claim any measurement accuracy Is it possible to revise these choices The measurement conditions reveal poor control of transitions from...

Page 19: ...T 4 Echo Alternating signal peak to peak 4 Volt max 0 65 V negative threshold voltage PT 0 0 Volt Ground PT 2 Synchro Trig on Positive edge 0 to 5 Volt PT 1 Selection windows Echo must be inside PT 3...

Page 20: ...NT 208D GB1 20 20 11 APPENDIX Calibration file example with a MiniSonic 2000 2 on a flow spool with wetted probes SM...

Reviews: