The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection scope to
verify fiber cleanliness. Plug the connectors on the Test Jumper into the OPM1 and LSSM. The OPM1 should
be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm button once to put it into
the dBm mode.
REFERENCING SETUP
OPM1
Page 4
M-UM038-01 RP 460 Reference and Test
DMT
Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection
scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355.
The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm
button once to put it into the dBm mode.
Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm
and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the
dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the
RP 460 back to the dBm mode.
Between -5.00 and -7.00dBm
Hold for 5 seconds
Reference is set
Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the
optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to
appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper
into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the
Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be
shown on the RP 460 screen.
Test Jumper
The LSSM output power is calibrated to -5dBm. If the OPM1 screen displays between -5.00dBm and
-7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the dB/dBm button.
The OPM1 screen will display 0.00dB, with some minor variation. DO NOT switch the OPM1 back to the
dBm mode.
SETTING REFERENCE
Page 4
M-UM038-01 RP 460 Reference and Test
DMT
Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection
scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355.
The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm
button once to put it into the dBm mode.
Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm
and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the
dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the
RP 460 back to the dBm mode.
Between -5.00 and -7.00dBm
Hold for 5 seconds
Reference is set
Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the
optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to
appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper
into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the
Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be
shown on the RP 460 screen.
Page 4
M-UM038-01 RP 460 R
efer
ence and T
est
DMT
Refer
encing Setup
The connectors on the T
est Jumper should be cleaned to appropriate standards. Use an i
nspection
scope to verify fiber cleanliness. Plug the connectors on the T
est Jumper into the RP 460 and DLS 355.
The RP 460 should be in the dBm measurement mode. If it is not
in the dBm mode, press the dB/dBm
button once to put it into the dBm mode.
Setting R
efer
ence
The DLS 355 output power is calibrated to -5dBm. If the RP 460
screen displays between -5.00dBm
and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. A
fter 5 seconds, release the
dB/dBm button. The RP 460 screen will display 0.00dB, with some
minor variation. DO NO
T switch the
RP 460 back to the dBm mode.
Betw
een -5.00 and -7.00dBm
Hold for 5 seconds
Refer
ence is set
Loss T
est
The T
est Jumper MUST remain attached to the DLS 355. Do not remove t
he T
est Jumper from the
optical link. The connectors on the T
est Jumper and the Fiber-Under-T
est should be cleaned to
appropriate standards. Use an inspection scope to verify fiber c
leanliness. Plug the T
est Jumper
into one end of the Fiber-Under-T
est using a bulkhead adapter as shown. Plug the other end of th
e
Fiber-Under-T
est into the L
C adapter on the RP 460. The Insertion Loss of the Fiber-Under-
Test will be
shown on the RP 460 screen.
OPM1
Between -5.00 and -7.00dBm
Page 4
M-UM038-01 RP 460 Reference and Test
DMT
Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection
scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355.
The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm
button once to put it into the dBm mode.
Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm
and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the
dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the
RP 460 back to the dBm mode.
Between -5.00 and -7.00dBm
Hold for 5 seconds
Reference is set
Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the
optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to
appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper
into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the
Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be
shown on the RP 460 screen.
Page 4
M-UM038-01 RP 460 Reference and Test
DMT
Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection
scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355.
The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm
button once to put it into the dBm mode.
Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm
and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the
dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the
RP 460 back to the dBm mode.
Between -5.00 and -7.00dBm
Hold for 5 seconds
Reference is set
Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the
optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to
appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper
into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the
Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be
shown on the RP 460 screen.
Hold for 5 seconds
Reference is set
Page 4
M-UM038-01 RP 460 Reference and Test
DMT
Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection
scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355.
The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm
button once to put it into the dBm mode.
Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm
and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the
dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the
RP 460 back to the dBm mode.
Between -5.00 and -7.00dBm
Hold for 5 seconds
Reference is set
Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the
optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to
appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper
into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the
Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be
shown on the RP 460 screen.
Test
Jumper
The Test Jumper MUST remain attached to the LSSM. Do not remove the Test Jumper from the optical link.
The connectors on the Test Jumper and the fiber-under-test should be cleaned to appropriate standards.
Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper into one end of the fiber-under-test
using a bulkhead adapter as shown. Plug the other end of the fiber-under-test into the LC adapter on the
OPM1. The Insertion Loss of the fiber-under-test will be shown on the OPM1 screen.
LOSS TEST
Fiber
-Under
-Test
Summary of Contents for TEK-3
Page 2: ...SECTION ONE LSMM LED LIGHT SOURCE QUICK START GUIDE NORTH AMERICA ...
Page 8: ...SECTION TWO LSMM LED LIGHT SOURCE TECHNICAL GUIDE NORTH AMERICA ...
Page 11: ...SECTION THREE OPM1 OPTICAL POWER METER QUICK START GUIDE NORTH AMERICA ...
Page 18: ...SECTION FOUR OPM1 OPTICAL POWER METER REFERENCE TEST GUIDE NORTH AMERICA ...