Arc calibration is a function to change the value of arc discharge voltage. It
is used for the computation program for splicing. In addition, the
calibration value of arc discharge cannot be changed in the splice mode.
For the purpose of Arc calibration, 12 core or 8 core single mode fibers are
used.
i.
Place the SM fiber to splicer.
ii. Click “OK”.
iii.
Result similar to following is displayed on the screen when measuring
is completed.
iv. The calibration can be stopped by pressing
even if it has not
completed.
The tip of the fiber should be removed after [arc calibration].
Otherwise, it may cause serious damage to the stripper blade.