Page 75
Warnings!
To avoid sever injuries from hot steam when opening the door:
It is strictly forbidden to lean on the autoclave.
It is strictly forbidden to place your hand or any part of your body
over the door.
Wear heat-resistant gloves or use the tray handle to remove the
load from the autoclave
To avoid severe injuries from hot steam when opening the door:
It is strictly forbidden to lean on the autoclave.
It is strictly forbidden to place your hand or any part of your body
over the door.
Since the autoclave is defined as a bio-hazard autoclave, a failed
cycle may leave un-sterilized waste in the chamber. Therefore
failed cycle will be followed by a slow exhaust and the door will
remain locked. To enable opening the door perform a new cycle.
The door can be opened only after a successful cycle.
4.
To release the door locking at the end of operation, press the UP key.
The same applies at power up after fail. Rotate the handle counter-
clockwise to pull out the locking arms handle from the retaining
brackets.
5.
Wear heat-resistant gloves or use the tray handle to remove the load
from the autoclave
6.
On completion of the cycle, the load shall be visual inspected to
ascertain that the load is dry, and that the colour of the sterilization
indicators turned to the required color.
7.
At the end of each working day close the main water valve.
10.11
Cycle by Clock mode
This mode enables the operator to define the time of the beginning of
the cycle. The maximum possible delay is 24 hours.
For more details, see sec. 6.8.1 "START CYCLE BY CLOCK"
Caution!
Before moving the autoclave, verify that the electrical, air and
water connections have been disconnected, and there is no
pressure in the chamber.