Page 43
5.15 Program 15: Isothermal
For liquids when the manufacturer recommends autoclaving at 60-
100
C (140-212
F) with no drying. Recommending for melting of agar,
pasteurization etc.
Caution!
This is not a sterilization program!
Put one PT100 inside the bottle, leave the second one hanging in the
chamber outside the bottle (see below).
Nominal parameters default settings
Heating temperature (default): 80
C (176
F).
Heating time: 20 minutes.
Operations sequence:
Heating:
Steam enters the chamber and heats it up until the Keep
Heat temperature is reached.
Keep Heat:
Keep-Heat temperature is maintained constant for the
preset Keep-Heat time.
Summary of Contents for 2840 ELG-D
Page 2: ...https tuttnauer com...
Page 19: ...Page 17 ON OFF On Off LAB048 0018 Near the Power switch...
Page 89: ...Page 87...