Page 51
7
Printer
7.1 Printer Output
The printing is on thermal paper with 24 characters per line and
contains the following information:
Date:
Time:
Ser. Num:
Model:
Version:
Cycle Num:
Cycle:
Ster Temp:
Ster Time:
When the sterilization cycle begins the printer starts printing the above
data.
After the preliminary printing, the autoclave starts performing the
sequence of operations of the cycle. The measured values of
temperature and pressure are printed at fixed time intervals, according
to the phase of the process, as shown in the table on the next page.
The data is printed from the bottom up, beginning with the date and
ending with "Cycle Ended". For an aborted cycle, "Cycle Failed" and
the Error message are printed (refer to "Displayed Error
Messages/Symbols").
For an example of a typical printout, see next page.
Note:
Please note that the print out goes from the bottom upwards.
Printer output
Description
Operator
:
____________
To be filled in manually by operator
Time:
12:14:47
Time sterilization cycle ended
Status: Cycle
Ended
00:24:43
101.3
099.7
Cycle finished time
D 00:31:23
101.9
023.4
The time, temperature and pressure during drying
D 00:28:23
106.2
109.7
The time, temperature and pressure during
exhaust
E 00:23:43
107.0
107.4
The time, temperature and pressure during
exhaust
E 00:22:08
134.5
311.9
The time, temperature and pressure during
exhaust
CLK 2:
12:12:10:00
Sterilization Process End time as registered by
Summary of Contents for 2840 ELG-D