Page 16
Trays are made of stainless steel.
Water reservoir is made of hard plastic material.
Door handle is made of hard plastic material, which is safe to touch
and thermo-insulated.
3.13 Safety Features
This autoclave includes built-in safety features such as:
Error message display.
Temperature dependent door locking system according to European
standards.
Electronic pressure and temperature measurement.
Safety relief valve to avoid build-up of excessive pressure.
Door switch enabling operation to be started only when the door is
closed.
Water level safety device.
Excess temperature protection.
3.14 Stickers Description
Symbol
Meaning
Part Number
Location
Caution! Hot
steam.
LAB048-0058
Near the safety
valve
Protective earth
(Ground)
LAB048-0020
Near the earthing
screw
Caution! Hot
surface.
LAB048-0023
On the rear cover
above the water
strainer
Press & Hold
Until “System
Ready” is
Displayed = 10
sec.
LAB048-0461
(Automatic
door only)
On the door right
edge
Don’t Touch
Microswitches
please Contact
Tuttnauer
Service
LAB048-0462
On the door frame
behind the door
cover
Don’t Touch
Microswitches
please Contact
Tuttnauer
Service
LAB048-0463
On the door frame
behind the door
cover