Page 40
Slow Exhaust:
Steam is exhausted from the chamber at a slow rate,
until it reaches the pressure of 30 kPa above the ambient pressure.
Then steam is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.
Note
: If the cycle fails, fast exhaust is performed through the bio-hazard
filter. Some water may remain in the chamber after a failed cycle. To
remove such water, press Start and repeat the cycle. If the problem
persists, call for service.
Drying:
N/A
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.12 Program 12: Vacuum Test (PV, pre-vacuum models
only)
Vacuum Test is a test program with the following parameters:
Nominal parameters default settings
Vacuum pressure: 17.0 kPa
Vacuum time 1: 5 minutes
Vacuum time 2: 10 minutes
Operations sequence:
Vacuum is produced in the chamber down to P1=17 kPa. At this
stage all the valves close. The autoclave remains in this stage for 5
minutes. This period enables the condition in the chamber to reach
equilibrium.
After the 5 minutes have elapsed, the printer records the pressure
that is referred to as P2. At this point the test begins and lasts 10
minutes.
At the end of the test, the printer records the results. The pressure at
the end of the test is referred to as P3. The rate of change of P3-P2
shall not exceed 0.13 kPa/min.).