Page 52
CLK 1:
12:12:10:00
two clocks
S 00:22:07
134.5
311.6
The time, temperature and pressure during
sterilization
S 00:22:06
134.5
311.6
The time, temperature and pressure during
sterilization
S 00:21:06
134.6
311.0
The time, temperature and pressure during
sterilization
S 00:20:06
134.5
310.1
The time, temperature and pressure during
sterilization
S 00:19:06
134.8
311.1
The time, temperature and pressure during
sterilization
S 00:18:06
134.5
315.8
The time, temperature and pressure during
sterilization
CLK 2:
12:08:08:00
Sterilization Process Start time as registered by
two clocks
CLK 1:
12:08:08:00
H 00:18:04
134.4
315.1
The time, temperature and pressure during
heating
H 00:16:35
128.9
268.4
The time, temperature and pressure during
heating
H 00:13:35
116.3
110.9
The time, temperature and pressure during
heating
A 00:08:48
089.6
077.1
The time, temperature and pressure during air
removal
A 00:05:48
091.3
049.4
The time, temperature and pressure during air
removal
W 00:03:22
092.5
015.2
The time, temperature and pressure during water
inlet
A 00:03:05
092.8
017.1
The time, temperature and pressure during air
removal
A 00:00:05
098.3
101.6
The time, temperature and pressure during air
removal
TIME
°C
kPa
End Temp:
120.0 °C
Ster Time:
7.0min
Sterilization time for selected program
Ster Temp:
134.0°C
Sterilization temperature in chamber for selected
program
Cycle:
Unwrapped
pouches
Cycle name
Cycle Num:
000082
Cycle number
Version:
1.0.00.00
Version A.B.CC.DD = 1.0.00.00
A: Door Type: Single Manual = 1
B: Vacuum Type = 0
C: Total number of Input/Output functionality that
are not as default = 00
D: Total number of parameters values that are not
as default = 00
Model:
2840 ELPV-D
Model name
Ser. Num:
000000000001 Model Serial number
Time:
11:50:05
Time sterilization cycle started