Miscellaneous Commands
∗
LRN?
Returns the complete set up of the instrument as a hexadecimal character
data block approximately 842 bytes long. To re-install the set up the block
should be returned to the instrument exactly as it is received. The syntax of
the response is LRN <Character data><rmt>. The settings in the instrument
are not affected by execution of the
∗
LRN? command.
LRN <character data>
Install data for a previous
∗
LRN? command.
∗
RST
Resets the instrument parameters to their default values (see DEFAULT
INSTRUMENT SETTINGS).
∗
RCL <nrf>
Recalls the instrument set up contained in store number <nrf>. Valid store
numbers are 0 - 9. Recalling store 0 sets all parameters to the default
settings (see DEFAULT INSTRUMENT SETTINGS).
∗
SAV <nrf>
Saves the complete instrument set up in the store number <nrf>. Valid store
numbers are 1 - 9.
∗
TRG
This command is the same as pressing the MAN/SYNC key. Its effect will
depend on the context in which it is asserted.
HOLD <cpd>
Set hold mode <ON> or <OFF>. This is the same as pressing the HOLD key.
FILTER <cpd>
Set the output filter to <AUTO>, <ELIP>, <BESS or <NONE>.
BEEPMODE <cpd>
Set beep mode to <ON>, <OFF>, <WARN>, or <ERROR>.
BEEP
Sound one beep.
LOCAL
Returns the instrument to local operation and unlocks the keyboard. Will not
function if LLO is in force.
Refer to Calibration section for remote calibration commands.
78