Chapter 5
54
Troubleshooting Table
Symptom
Possible Cause
Corrective Action
Erratic zero reading.
Leak.
Check all connections
for leaks.
Carefully tighten sample
inlet screws (
DO NOT
over-tighten!).
If using an impactor,
remove, clean and re-oil
the impactor disc. Make
sure the O-ring is in place
and not damaged.
Make sure the impactor
disc is seated squarely
into the fitting, before
replacing the inlet (see
“
”, in
Chapter 4, for more
information).
Leak through
inlet gasket.
Cover the inlet with
thumb. If there is a leak
the pump will continue
to run.
Remove the inlet and
cover the opening with
your thumb to block the
air flow. If the error ‘Flow
Blocked’ is displayed,
replace the inlet gasket
and re-assemble.
Dirty inlet and/or
sample tube.
Clean inlet. Clean or
replace tubing.
Extreme temperature
swings during
operation.
Zero instrument, at
ambient temperature,
before beginning test.
Protect from extreme
temperature swings.
Internal optics
chamber is
contaminated.
Return to factory for
cleaning and servicing.
Nothing visible on
display.
Unit not switched on.
Press the
ESC
key firmly
for 2 seconds minimum.
Low or dead battery.
Replace the battery or
plug in the AC adapter.
Summary of Contents for SidePak AM520
Page 8: ...viii This page intentionally left blank...
Page 22: ...6 Chapter 1 This page intentionally left blank...
Page 31: ...Operation 15 Battery Screws Remove these two screws to remove the battery...
Page 56: ...Chapter 3 40 This page intentionally left blank...
Page 78: ...Chapter 5 62 This page intentionally left blank...
Page 88: ...Appendix B 72 This page intentionally left blank...
Page 90: ...Appendix C 74 This page intentionally left blank...
Page 91: ...75 Appendix D CSA Certificate of Compliance Certificates of Compliance on file at TSI...
Page 92: ...Appendix D 76 AM520 Declaration of Conformity...
Page 93: ...CSA Certificate of Compliance 77...
Page 94: ...Appendix D 78 AM520i Declaration of Conformity...
Page 95: ...CSA Certificate of Compliance 79...
Page 96: ...Appendix D 80 This page intentionally left blank...
Page 97: ......