Chapter 4
50
Cleaning the Diesel Particulate Matter DPM Cyclone
The DPM Cyclone should be cleaned prior to each use. In most cases,
simply cleaning the grit potentiometer is all that is needed. Inspect the
inside of the cyclone body regularly and clean it if necessary.
1. Unscrew the grit potentiometer from the bottom of the cyclone.
Remove the stainless steel cyclone clip and pull the cap off.
2. Hold the open end of the grit potentiometer down and tap it on a
hard surface to dislodge particles. Repeat with the cyclone body.
N O T I C E
If dirt is visible inside either the grit potentiometer or the cyclone
body, it may be necessary to blow canned/pressurized air into the
cyclone parts and/or to clean them with soap and water. A mild
solvent like isopropanol may also be used inside the cyclone. Make
certain that the cyclone is perfectly dry before re-assembly and use.
3. Re-assemble the cyclone.
Note that the stainless steel
cyclone clip that holds the cap
onto the body will only fit
one way.
The cyclone cleaning procedure
is now completed.
Exploded View of
0.8 µm DPM Cyclone
Summary of Contents for SidePak AM520
Page 8: ...viii This page intentionally left blank...
Page 22: ...6 Chapter 1 This page intentionally left blank...
Page 31: ...Operation 15 Battery Screws Remove these two screws to remove the battery...
Page 56: ...Chapter 3 40 This page intentionally left blank...
Page 78: ...Chapter 5 62 This page intentionally left blank...
Page 88: ...Appendix B 72 This page intentionally left blank...
Page 90: ...Appendix C 74 This page intentionally left blank...
Page 91: ...75 Appendix D CSA Certificate of Compliance Certificates of Compliance on file at TSI...
Page 92: ...Appendix D 76 AM520 Declaration of Conformity...
Page 93: ...CSA Certificate of Compliance 77...
Page 94: ...Appendix D 78 AM520i Declaration of Conformity...
Page 95: ...CSA Certificate of Compliance 79...
Page 96: ...Appendix D 80 This page intentionally left blank...
Page 97: ......