Sensitivity
Sensor type
90
light scattering, 650 nm laser diode
Aerosol concentration
range
0.001 to 100 mg/m
3
(calibrated to
respirable fraction of ISO 12103-1,
A1 Test Dust)
Particle size range
0.1 to 10 µm
Minimum resolution
0.001 mg/m
3
Zero stability
±0.001 mg/m
3
over 24 hours using
10-second Time Constant
Temperature coefficient
approxi0.0005 mg/m
3
per
C
(for variations from temperature at
which instrument was last zeroed)
Flow Rate
Range
User-adjustable, 0 to 1.8 L/min
N O T I C E
Upper range is dependent on
impactor or cyclone attached to the
AM520/AM520i.
Temperature Range
Operating range
0
C to 50
C (32
F to 120
F)
Storage range
-20
C to 60
C (-4
F to 140
F)
Operational humidity
0 to 95% RH, non-condensing
Time Constant
(OLED display)
Range
User-adjustable, 1 to 60 seconds
Summary of Contents for SidePak AM520
Page 8: ...viii This page intentionally left blank...
Page 22: ...6 Chapter 1 This page intentionally left blank...
Page 31: ...Operation 15 Battery Screws Remove these two screws to remove the battery...
Page 56: ...Chapter 3 40 This page intentionally left blank...
Page 78: ...Chapter 5 62 This page intentionally left blank...
Page 88: ...Appendix B 72 This page intentionally left blank...
Page 90: ...Appendix C 74 This page intentionally left blank...
Page 91: ...75 Appendix D CSA Certificate of Compliance Certificates of Compliance on file at TSI...
Page 92: ...Appendix D 76 AM520 Declaration of Conformity...
Page 93: ...CSA Certificate of Compliance 77...
Page 94: ...Appendix D 78 AM520i Declaration of Conformity...
Page 95: ...CSA Certificate of Compliance 79...
Page 96: ...Appendix D 80 This page intentionally left blank...
Page 97: ......