Troubleshooting
55
Symptom
Possible Cause
Corrective Action
Mass concentration
number is blinking
and stays on the
same value.
Instrument has
reached the
calibrated limit of its
range (multiplied by
Cal Factor).
N/A
No keypad response.
Display shows a red
lock icon.
Keypad is locked out.
When the keypad is
locked, the display
shows a red lock icon.
To unlock the keypad,
press and hold the ▲ key
and press
.
See
Chapter 3 for more
information.
The battery icon is
flashing red.
Low battery charge.
Recharge battery or use
AC adapter.
Green LED on battery
flashing on for
3 seconds and off for
1 second.
The battery is either
too hot or too cold
to charge.
Allow battery to warm or
cool to room temperature.
Green LED on battery
flashing rapidly.
Internal battery pack
fault. Battery cannot
charge.
Replace battery.
FLOW BLOCKED
message is displayed.
Inlet flow is blocked.
Remove obstructions.
Check for pinched sample
tube. Check for correct
installation of impactor
inlet (if used).
Outlet flow is blocked.
Remove obstructions from
outlet flow fitting (next to
connectors).
Internal screen filter
is plugged.
Return to factory for
servicing, or replace
internal screen filter. See
“
” section, in
Chapter 4, for more
information.
Summary of Contents for SidePak AM520
Page 8: ...viii This page intentionally left blank...
Page 22: ...6 Chapter 1 This page intentionally left blank...
Page 31: ...Operation 15 Battery Screws Remove these two screws to remove the battery...
Page 56: ...Chapter 3 40 This page intentionally left blank...
Page 78: ...Chapter 5 62 This page intentionally left blank...
Page 88: ...Appendix B 72 This page intentionally left blank...
Page 90: ...Appendix C 74 This page intentionally left blank...
Page 91: ...75 Appendix D CSA Certificate of Compliance Certificates of Compliance on file at TSI...
Page 92: ...Appendix D 76 AM520 Declaration of Conformity...
Page 93: ...CSA Certificate of Compliance 77...
Page 94: ...Appendix D 78 AM520i Declaration of Conformity...
Page 95: ...CSA Certificate of Compliance 79...
Page 96: ...Appendix D 80 This page intentionally left blank...
Page 97: ......