6-2
AeroTrak
®
Portable Airborne Particle Counter
Symptom
Possible Cause
Corrective Action
Battery does not charge
The unit must be turned on but not in
sampling mode for the battery to charge.
Turn on unit. Green LED by on/off button
should be lit.
Unit is actively taking data.
The unit will not recharge the battery
while taking a measurement. Once the
measurement is complete, the unit will
begin recharging the battery.
Unit not put in standby mode when turned
off.
Select Standby/Charge when shutting off
the instrument if you want the battery to
be charged.
BATTERY ERROR
Low battery (<10%).
Recharge battery or connect AC cord.
Single battery in 100 L/min Model 9500, or
two batteries where one has been only
partially charged
The 100 L/min Model 9500 requires two
batteries (within 25% charge of each
other) to take a sample. If only 1 battery
is installed, then the unit will not start a
measurement and will display the battery
warning. To correct, place a second
charged battery in the instrument.
FLOW ERROR
Instrument was unable to control flow rate
[i.e., if sample tubing used is too long
(greater than 12 m on 28.3 and 50 L/min
instruments, 8 m on 100 L/min instrument)].
Restart measurement.
Pressure drop across inlet may be too
large.
Lower pressure drop across inlet by
using larger diameter tubing, less tubing,
and/or adding a bleed valve.
Inlet tubing kinked.
Straighten tubing.
Inlet not at ambient pressure.
Do
not
subject the unit to other than
ambient pressure conditions.
LASER POWER /
DETECTOR WARNING
Excessive direct light is entering the
aerosol inlet.
Remove instrument from direct light.
Optical path blocked.
Return to factory for service.
Nozzle is misaligned. Fiber attached on the
nozzle tip.
Contact TSI and return to factory.
Detector board damaged. Laser power is
normal.
Return to factory or factory authorized
service centers for service.
Summary of Contents for AEROTRAK 9310
Page 2: ......
Page 18: ...1 6 AeroTrak Portable Airborne Particle Counter This page intentionally left blank ...
Page 76: ...5 2 AeroTrak Portable Airborne Particle Counter This page intentionally left blank ...
Page 82: ...7 4 AeroTrak Portable Airborne Particle Counter This page intentionally left blank ...
Page 90: ......
Page 91: ......