2-8
AeroTrak
®
Portable Airborne Particle Counter
The printer has a feed
and stop
button as well as an LED
indicating that the printer is ready. The feed button can be held down at
the end of a print to allow enough space to tear off the paper. If you
unintentionally start a print (especially something very long), stop printing
with the stop
button.
U s i n g P e r i p h e r a l s
Connecting to the Alarm Closure
This connector provides two pins for a contact closure to control an
external alarm. The contact closure is normally open and rated for 0 to
60 V AC/DC at 1.5A peak, 0.5A continuous.
The “alarm out” connection is
rated for 60 VDC insulation. The contact is closed under alarm conditions
determined by the programming of the device. When used with an
externally powered device, this can trigger a visual or audible local alarm
(such as a light pole).
W A R N I N G
The monitor Alarm Output function should
not
be used to detect
hazardous conditions or to provide an alarm for protecting human life,
health, or safety.
C a u t i o n
The alarm switch must not be wired to AC power! Failure to properly
install the user alarm could damage the instrument and/or void the
instrument warranty! Please read and follow all instructions before wiring
or operating the user alarm.
Alarm
Filter
Scanner
Probe
USB A
(Flash Drive)
Summary of Contents for AEROTRAK 9310
Page 2: ......
Page 18: ...1 6 AeroTrak Portable Airborne Particle Counter This page intentionally left blank ...
Page 76: ...5 2 AeroTrak Portable Airborne Particle Counter This page intentionally left blank ...
Page 82: ...7 4 AeroTrak Portable Airborne Particle Counter This page intentionally left blank ...
Page 90: ......
Page 91: ......