Model 3450
5–15
PRECISION
The gauge can produce test results with precision limits as low as
0.1 pcf (1.6 kg/m
3
) under certain conditions. The
Precision
function
is used to determine the count time required to achieve the precision
entered by the operator. If the precision requested is within range
and the time required to calculate the required result does not
exceed 60 minutes, the following procedure can usually produce the
required results.
To execute the
Precision
function, first ensure that the gauge is in
the correct mode:
Soil
or
Asphalt
. Press
〈
7
〉
from the
Setup
menu
shown on page 5–3. The gauge displays:
Enter the desired precision and press
〈
ENTER/START
〉
. The gauge
displays:
If the gauge is in the correct position, press
〈
ENTER/START
〉
. The
gauge displays:
5. SET
UP M
EN
U
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...