![Troxler 3430 Plus Manual Of Operation And Instruction Download Page 51](http://html1.mh-extra.com/html/troxler/3430-plus/3430-plus_manual-of-operation-and-instruction_1172534051.webp)
Model 3430 Plus & 3440 Plus
4–5
If the standard count
fails
and it has been
more than
a month
since the last standard count, then accept the standard count by
pressing the
〈
YES
〉
key. At the
Do You Want To Erase Last
Four Standard Counts?
prompt, press the
〈
YES
〉
key. The
gauge displays:
Follow the displayed instructions, and take four additional
standard counts. The last of these counts will be compared to
the previous four, and the standard counts should pass. If not,
repeat the procedure. If it still fails, perform a stat test (see page
5–5) and a drift test (see page 5–7), then contact your Troxler
representative.
If the standard count
fails
and it has been
less than
a month
since the last standard count, but the count was performed
correctly and the failure is less than 5%, press the
〈
YES
〉
key. If
the failure is more than 5%, press the
〈
NO
〉
key. Ensure that the
gauge is properly positioned on the reference standard block
(Figure 4–1). Verify that the standard count site meets the
criteria listed on page 4–2. Take another standard count and
accept it if it fails by less than 5%. If the standard count fails
four times, perform a stat test (see page 5–5) and a drift test (see
page 5–7), then contact your Troxler representative.
After the standard count passes, the operator can take measurements
with the gauge. When not taking readings, always keep the source
rod in the
SAFE
position. For added user safety, the source rod
automatically retracts to the
SAFE
position when the gauge is lifted
using the handle.
4. U
SIN
G T
HE
G
AU
GE
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...