7–3
DENSITY OFFSET
The density offset is the difference between the gauge reading and
the density as measured by an alternate method. Wet density values
should always be used for this comparison.
To access the
Density Offset
menu, press
〈
1
〉
at the
Offset
menu
shown above. The gauge displays:
The gauge displays the current density offset on the second line.
To enable the displayed density offset, press
〈
1
〉
. The gauge
enables the offset, displays a brief confirmation message, and
returns to the
Ready
screen.
To disable the density offset, press
〈
2
〉
. The gauge disables the
offset, displays a brief confirmation message, and returns to the
Ready
screen.
NOTE
The density offset is also disabled when the gauge is
turned off.
To enter a new density offset, press
〈
3
〉
. The gauge prompts for
the density offset as shown:
To change the offset, select the offset sign (positive or
negative), enter the density offset, and press
〈
ENTER/START
〉
.
The gauge enables the new density offset and returns to the
7. C
AL
IBR
AT
IO
N O
FF
SET
S
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...