Model 3430 Plus & 3440 Plus
Appendix C–7
Table C–1. Error Messages (Continued)
Error Message
Possible Causes
Solution
Error Writing to
the File
An error occurred when
attempting to store data to a
USB flash drive.
Try to store the data again. If
error recurs, try a different
USB flash drive.
Error! Stat Test
Not Valid
Operator has tried to run a
drift test without having a valid
stat test.
Perform a stat test as
described on page5–5, then
run a drift test.
Gauge Requires
Re-calibration!
Contact Tech
Support
The calibration constants
stored in the gauge have
become corrupted.
Contact the nearest Troxler
Service Center (see page iii).
Gmb Target
Exceeds Limits!
The target value must be
between 20.0 and 200.0 pcf.
The operator has entered a
value outside this range.
Enter a target value between
20.0 and 200.0 pcf.
Gmm Target
Exceeds Limits!
The target value must be
between 20.0 and 200.0 pcf.
The operator has entered a
value outside this range.
Enter a target value between
20.0 and 200.0 pcf.
Invalid Access
Code
Operator entered an invalid
access code.
Enter the correct access code
(see page xiii).
Invalid Depth
The source rod is not at a
valid depth when starting a
measurement.
Ensure the source rod
position is at a valid depth
when starting a measurement.
Contact the nearest Troxler
Service Center (see page iii).
Low Battery.
Gauge Shutting
Down. Turn off &
Charge
The following three items are
all true
: the charger is not
connected, the NiMH batteries
are low, and the alkaline
batteries need replacing.
Charge the NiMH batteries
and replace the alkaline
batteries.
Moisture Offset
Exceeds Limits!
The moisture offset (k) value
must be between –200 and
200. When performing a
moisture offset, the resulting k
value was outside of these
limits.
Enter a new moisture offset as
described on page 7–4.
No active project!
Operator tried to store
measurement data, but no
project is active.
Select an existing project or
create a new project and
select as active as described
in Chapter 8.
Proctor Target
Exceeds Limits!
The target value must be
between 20.0 and 200.0 pcf.
The operator has entered a
value outside this range.
Enter a target value between
20.0 and 200.0 pcf.
C. T
RO
UB
LE
SH
OO
TIN
G
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...