Model 3400-B Series
5-3
Sample
No.
%M Oven
Dry
%M Gauge
%Oven Dry -
% Gauge
1
2
3
4
5
4.5
4.0
7.2
6.7
5.1
8.6
5.8
9.7
8.6
7.3
-4.1
-1.8
-2.5
-1.9
-2.2
Average
5.5
8.0
-2.5
Table 1. Sample Moisture Correction Chart
The difference between the oven dry and the gauge samples is
expected to vary from sample to sample due to normal
variation. If the difference indicates the gauge is sometimes
higher and sometimes lower than the oven dry, no correction
may be needed. If the difference indicates the gauge reads
consistently
higher or lower than the oven dry, a correction
factor is needed.
To determine the correction factor, calculate the average value
and proceed as follows:
FOR THE 3411-B
Calculate the correction factor using the average value.
%M Oven Dry - %M Gauge
100 + %M Gauge
x 1000
Dial this value into the moisture correction switches on the
3411-B scaler, paying attention to the algebraic sign. In the
example above, the moisture correction is:
-2.5
100 + 8.0
x 1000
= -23
Summary of Contents for 3411-B
Page 3: ...Model 3400 B Series iii...
Page 10: ......
Page 15: ...Model 3400 B Series 1 3 Figure 1 Gauge and Accessories...
Page 20: ...2 4 Figure 2 3411 B Front Panel Controls...
Page 30: ...3 8 NOTES...
Page 44: ...5 8 Figure 4 Roller Pattern...
Page 76: ...Appendix A 16 NOTES...
Page 91: ...Model 3400 B Series Appendix C 3 Figure 15 Direct Transmission and Backscatter Geometries...
Page 92: ...Appendix C 4 Figure 16 Backscatter Surface Density Effects...
Page 104: ......
Page 108: ...Appendix E 4 NOTES...