Model 3400-B Series
3-5
NOTE
On soils that have not previously been checked by a
nuclear gauge, a moisture sample may be obtained
at a point of test to determine if a correction factor is
needed. If a correction factor is necessary, it can be
determined by following the procedures starting on
page 5-2. A correction factor for a given soil need
only be determined once. The correction can then be
reused in future testing with the same gauge.
1.
Press the
〈
MEASURE
〉
key. The display will show the gauge
counting and ERR in the upper-left corner. After one minute
the counting will stop.
2.
Press
〈
WD
〉
and record the Wet Density (kg per cubic meter
or PCF).
3.
Press
〈
M
〉
and record the Moisture.
4.
Press
〈
DD
〉
and record the Dry Density (WD-M).
5.
Press
〈
%M
〉
and record percent moisture (M divided by DD)
Summary of Contents for 3411-B