5-6
ON THE 3401
Subtract the offset value from the moisture measurement count
before you divide by the moisture standard count.
ON THE 3411
Set this value on the MOISTURE CORRECTION switches. Press
and hold the
〈
SHIFT
〉
key while depressing
〈
MC
〉
. The value set
on the MOISTURE CORRECTION switches should now appear on
the display. This value is preset negative so the + switch need
not be moved.
The entered value for moisture offset will be retained by the
gauge until a) it is changed by entry of a new offset, b) a new set
of standard counts is taken, or c) the gauge is turned off.
Be sure to return the moisture correction switches to 00 or to
the values representing the proper moisture correction factor
after entering the trench correction value.
The moisture offset value can be removed by setting the
MOISTURE CORRECTION switches to 00 and pressing
〈
SHIFT
〉
and
〈
MC
〉
.
Summary of Contents for 3411-B