Direct transmission wet density will not be affected by a trench
wall or vertical structure. Moisture and dry density (WD-M)
may be affected when a gauge is operated within 0.6 meters (24
inches) of a large vertical structure with a moisture content of
240 kg/m
3
(15 pcf) or greater.
Fore trenches with a width of one meter (40 inches) or more,
the moisture error will be less than 24 kg/m
3
(1.5 pcf) when the
measurement is performed in the center of the trench.
Correction Procedure
1.
Obtain the daily standard counts following the steps outline
on page 3-2. (If a valid set of standard counts has already
been obtained, it may be used.) Note that the standard
counts are taken at a site
away
from the trench.
2.
Enter the trench and take a set of counts (not a standard
count) with the gauge
placed on the reference standard
at the
desired test site. The distance from the wall must be the
same as the area to be tested. This count may be taken with
the PWR/TIME switch set on NORM. On the 3411, do not
press
〈
SHIFT
〉
because your standard count will be erased.
3.
Determine the moisture offset value by subtracting the
moisture standard count from the second moisture count
obtained in the trench. Record this number for future tests
performed at this site and with this gauge at the same
distance from the wall.
Summary of Contents for 3411-B
Page 3: ...Model 3400 B Series iii...
Page 10: ......
Page 15: ...Model 3400 B Series 1 3 Figure 1 Gauge and Accessories...
Page 20: ...2 4 Figure 2 3411 B Front Panel Controls...
Page 30: ...3 8 NOTES...
Page 44: ...5 8 Figure 4 Roller Pattern...
Page 76: ...Appendix A 16 NOTES...
Page 91: ...Model 3400 B Series Appendix C 3 Figure 15 Direct Transmission and Backscatter Geometries...
Page 92: ...Appendix C 4 Figure 16 Backscatter Surface Density Effects...
Page 104: ......
Page 108: ...Appendix E 4 NOTES...