14
11
The Elapsed Time Meter also
indicates the source of the problem
should the unit malfunction. See the
Troubleshooting Section on page 20
for details.
period, it will go into alarm
again, the LED will turn solid red,
and the alarm will sound.
After 14 months of lamp
operation, the alarm postpone will
not work until the lamp is replaced
and the time meter is reset.
Elapsed Time Meter Re-Set
Function:
After the lamp has been changed,
the Elapsed Time Meter must be
reset by following the
procedure below:
a) Disconnect the power supply.
Ensure the ballast is left
unplugged for 10 seconds.
b) Press and HOLD the reset
button.
c) Connect the power supply to
the outlet while continuing to
press the reset button. The
display and LED will flash for
about 3 seconds.
d) Continue to hold the reset
button until the LED flashes
red and the display flashes
“00”.
Dry Contacts
This feature provides switching for
the operation of a solenoid valve
and/or remote alarm. When the
lamp is not operating properly or
the UV sensor signals that the UV
output is low, the contacts will
open causing the solenoid to stop
the water flow and/or a remote
alarm to sound. The dry contact
remains open if the lamp has been
in operation for 12 months or
more.
Remote Options Cord
must be ordered to use Dry
Contacts.
UV Sensor
The UV sensor measures the
amount of UV light reaching it,
allowing the system to monitor
whether the intensity is above the
minimum required for proper dis-
infection. The sensor is factory
calibrated and is not field
adjustable. See your dealer for
proper system sizing.
Alarm Postpone Button
The alarm postpone button has two
functions.
24-Hour Postpone Function:
When the unit is in alarm, the LED is
red and an alarm sounds. If you press
the Alarm Postpone Button for less
than two seconds, the LED will flash
red and the audible alarm will stop.
The unit is still in alarm, but the
audio alarm stops for your convenience
until you can contact a dealer.
This alarm will re-occur after 24 hours
if its cause has not been corrected.
If the unit detects another problem
during the 24-hour alarm postpone
6. Reconnect the power supply and
check that the LED is green. The
alarm may sound until the lamp
operates properly.
7. Open the water line, and check
for leaks.
8. Clean the distribution lines - refer
to item
#
11 on page 8.
Note:
During lamp replacement, the
ground and strain relief wire of the
lamp harness should always remain
connected to the stud on the reactor.
Wires are fastened to the stud by a
lock nut.
Caution:
The distribution lines must be cleaned
following any maintenance procedure
in which the water in the lines may
have been exposed to the air or to any
undisinfected water. The distribution
lines must also be cleaned whenever
the system loses power and water
use continues.