Getting Started
SeaKing & SeaPrince Imaging Sonars
0374-SOM-00001, Issue: 08
8
© Tritech International Ltd.
The Node table should display the Node of the connected Sonar (normally Node 2, but the
label on the unit should state what it has been set to) There may also be additional nodes,
depending on how the system is all connected. A SCU or SeaHub will appear as Node 252,
whereas a MicronNav Hub will appear as Node 90.
The Status of the Sonar Node should be OK.
If the Status is UNKNOWN check the following: For RS232 wiring Pin 5 is properly connected
For all wiring The main comms pins (1 and 2 on the waterblock) are wired correctly and have
not been crossed
With the Status showing as OK, select the Applications->Single Sonar option
The data screen will report a centering message, as part of the Sonar boot up, then will start
to scan.
It is recommended that the Super SeaKing Sonar is only run in air for short periods (1 hour
maximum) as the unit will become hot to the touch without a body of water to help dissipate
the heat.