Maintenance
SeaKing & SeaPrince Imaging Sonars
0374-SOM-00001, Issue: 08
36
© Tritech International Ltd.
6. On the top most PCB, locate the two hexagonal pictures on the PCB
7. Using the 5.5mm spanner secure the retaining nuts at these locations
8. Using the short nose flush pliers, fit the bullet post at the remaining points
9. Carefully insert the electronics block back into the Body tube, ensuring that the dowel pin
on the connector endcap lines up with the recess point on the Body tube
10.Fit the connector endcap locking ring
11.Using the M3 Allen key, fit the MAIN and AUX (if applicable) waterblocks. Each of the
waterblock screws should be tightened, in turn, by a quarter turn until the waterblock is
flush and level to the connector endcap
12.Using the M3 Allen Key, fit the Pressurised blanking cap (if applicable) either directly onto
the AUX port, or on the AUX waterblock if it not to be used.
13.Visually inspect the unit to ensure that all surfaces have mated correctly
6.4. Super SeaPrince regular maintenance
General Guidelines
Caution
It is essential to have a regular maintenance schedule so that any defects arising
from corrosion or erosion can be spotted early and corrected before they cause
severe damage to the unit. It is recommended that the unit is annually serviced
and can be returned to
Tritech International Ltd
for this purpose
Competent, trained, personnel can perform regular preventative maintenance on these units.
Contact
Tritech International Ltd
for more details on the training courses available.
Servicable Items
The standard Super SeaPrince has very few user servicable parts.
Item Part
Number
Description
1 S00009
0161-16 O-ring
2 S00987
Waterblock
3 S01182
ASSY
Pressurised
blanking cap
The o-ring seals should be regularly inspected, cleaned and lubricates with the appropriate
greasing compound. The body of the unit should also be inspected for any obvious signs of
corrosion, especially in mating surfaces (such as the waterblock).