SafeTcheck OPeRaTOR'S MaNUaL
26
Operating Instructions
Overview
All Data Logging Utilities, display adjustment, measured values, earth current options (when fitted) and the setting of time & date
are available from this menu.
The Utilities Menu is accessed via the F2 key when the
SafeTcheck
is at the Start Prompt. Once in the menu, the user may scroll
though the list by using the F1 and F2 keys.
To select an option, use the scroll keys. Once highlighted, press "Start" to select the option or "Reset" to exit. Follow screen
messages to select, then enable or disable an option. A will appear at the right of an enabled option.
Menu Options
Upload test results
(See Data Logger Operating Instructions)
Delete test results
(See Data Logger Operating Instructions)
Enable database
(See Data Logger Operating Instructions)
Ask for Venue
(See Data Logger Operating Instructions)
Ask for Technician
(See Data Logger Operating Instructions)
Ask for App#
(See Data Logger Operating Instructions)
Ask for Visual P/F
(See Data Logger Operating Instructions)
Ask for Tag#
(See Data Logger Operating Instructions)
Ask for Comment
(See Data Logger Operating Instructions)
Earth continuity 0.2A
(When High Current Earth Test option is fitted) The Earth Continuity Current maybe set
to either 0.2A as a default or as a 'one off'
Routine test 10A
(When High Current Earth Test option is fitted) The Earth Continuity Current maybe set
to either 10A as a default or as a 'one off'
Type test 25A
(When High Current Earth Test option is fitted) The Earth Continuity Current maybe set
to either 25A as a default or as a 'one off'
Measured values
If enabled, Measured values will be displayed for failed tests
Adjust display
Adjust the contrast of the LCD display
Set Time/Date
(See Data Logger Operating Instructions)
Switch between
and Probe Mode. (see this section for more details).
Fail on poor cont.
Fail on poor continuity – if set to 'off', will allow the operator to bypass the check for a closed mains
switch (it is recommended that this is always set to 'on')
Printer Enable
If enabled, tags may be printed on completion of a test (is disabled if 'Enable database' is off)
Earth Threshold
Sets the earth resistance threshold to 0.2, 0.5 or 1
Ω
(1.0 Ohms is the default for AS/NZS 3760).
(See warning)
Set retest period
Sets the 'Default retest period' (1, 3 & 6 months and 1, 2 & 5 years); 'Prompt every test'; or 'No
retest period'
utilities Menu
(F2 key)