TRIO TeST & MeaSUReMeNT
19
Copyright 2015 all rights reserved
1 Ensure the appliance is switched off.
2 Plug the appliance into the separate Run Test socket.
3 Connect the return lead to the
SafeTcheck
via the
Earth Return Socket.
4 Connect the Clamp of the Return Lead to a convenient
earthed metal point on the appliance under test.
WaRNINg
Upon completion of a safety test and subsequent
removal of test lead, please ensure all switches, speed
or thermostat controls are returned to their ‘OFF’ or
minimum position.
5 Switch the appliance on.
6 Select Class 1 Run Test under Other
APPLIANCE menu (F1 and F2 keys move the
selection of mode up or down respectively)
7 Press START button and follow the screen
instructions. The appliance will operate for
up to 20 seconds during the run test.
8 For Class I appliances requiring a longer than
normal run test duration, move to Long Class I
Run under OTHER APPLIANCE menu.
Note 1:
Pressing RESET at any point during the Run
Test will shorten the test duration. This does not invalidate
the test result.
Note 2:
If Data Logging is switched on then additional
information will be requested. See section entitled
”Data Logging”
WaRNINg
The appliance will operate for up to 20 seconds in the
run test mode. Ensure rotating parts or heating elements
are able to operate safely. Appliances such as drills, angle
grinders etc should be firmly restrained during the run test.
Run testing earthed (class 1) appliances
SWITCh APPLIANCE ON & SET
APPLIANCE ThERMOSTAT OR
SPEED CONTROL TO MAxIMuM