www.trilithic.com
www.trilithic.com
Page 23
8310 RSA Operation Manual
Return Sweep Analyzer
Setting Up the Active Channel Plan
From the
SETUP MENU
, highlight
PLAN SETUP
and then select the
Enter
button to display the
PLAN SETUP MENU
.
Use the arrow buttons to highlight the desired
channel plan setting, and then select the
Enter
button to adjust it.
The 8310 RSA has a minimum sweep frequency of 5.0625 MHz
and a maximum of 65.25 MHz, with a minimum step frequency
of 93.75 kHz.
Setting Start, Stop, & Step Frequency
To adjust the start, stop, and step frequency of
the sweep, perform the following steps:
1. From the
PLAN SETUP MENU
, use the
arrow buttons to highlight
START
,
STOP
,
or
STEP,
and then select the
Enter
button
to edit the frequency.
2. Use the up/down arrow buttons to adjust
the value of the selected digit and then
use the left/right arrow buttons to change
between the digits. Once you have
finished editing the frequency, select the
Enter