CODE/NAME 

DESCRIPTION 

SELECTION LIST 

F9 

Display 
Divisions

 

Sets the interval value. Use together with F10.  

1

 


F10 

Decimal Pt. 

Sets the decimal point value. Use together with F9. 

0

               0.0 

0.00        0.000 
0.0000         00 

F16 

Zero Calibra-
tion 

Places  indicator  into  live  zero-calibration  mode.  Scrolling  down  with 
the ZERO key one level begins the procedure. 

Press 

ZERO

 key 

to begin sequence 

F17 

Span Calibra-
tion

 

Places  indicator  into  live  span  calibration  mode.  Scrolling  down  with 
the ZERO key one level begins the procedure. 

Press 

ZERO

 key 

to begin sequence 

F18 

View Calibra-
tion Data 

Activates  the  function  that  allows  you  to  view  calibration  values. 
Scrolling down with the ZERO key one level begins the procedure.  

Press 

ZERO

 key 

to begin sequence 

F19 

Key-in Zero 

Allows  you  to  key-in  a  known  zero  calibration  value.  Scrolling  down 
with the ZERO key one level begins the procedure. 

Press 

ZERO

 key 

to begin sequence 

F20 

Key-in Span 

Allows  you  to  key-in  known  span  calibration  values.  Scrolling  down 
with the ZERO key one level begins the procedure. 

Press 

ZERO

 key 

to begin sequence 

F21 

Factory Reset 

This sub-

menu will reset all parameters in the “F” and “A” menu to the 

default  settings.  It  will not  overwrite  any  previously  saved  calibration 
data. 

USE WITH CAUTION!

 

Press the 

ZERO

 

key twice to exe-
cute

 

F24 

Piece Count 
Enable 

Used to enable or disable the piece count feature. This feature must 
be disabled for commercial applications. 

"0" = Disable piece count                   "1" = Enable piece count 

0

 

 
 

Entering the 

User (“A”) 

 

 

1.  Switch off the indicator 

2.  Locate  the  slide  switch  on  the  rear  cover  and  move  it  to  the  opposite  position. 

NOTE:

 A metal plate held on by two drilled-head screws conceals the slide switch. 

3.  Switch  on  the  indicator.  The  display  shows 

“F   1” to  indicate that you  are in Setup Menu 

mode. 

4.  Use the TARE (left) or PRINT (right) keys 

to move right or left in the Setup (“F”) menu until 

the indicator shows 

“A   1”. 

5. 

Move from one “A” parameter to the next by using the TARE (left) or PRINT (right) keys. For 
example, to go from  A1 to A2, press the  PRINT key. To go from A2 back to A1, press the 
TARE key. 

6.  Once  you  have  arrived  at  the  proper 

“A”  menu  parameter,  e.g.  “A  1”,  press  the  

ZERO (down) key once to arrive at the selection level. The indicator displays the current pa-
rameter setting. 

Summary of Contents for TI-500E

Page 1: ...MODEL TI 500E Digital Weight Indicator Manual Version 1 22S June 28 2022...

Page 2: ...ject to change without notice Contents DISPLAY KEYPAD DETAILS 2 INSTALLATION OVERVIEW 5 INSTALLATION 5 GETTING STARTED 6 INDICATOR CONFIGURATION 7 CONFIGURATION MENUS 7 ENTERING THE SETUP F CONFIGURAT...

Page 3: ...ht mode NET The indicator is in Net Weight mode TARE Indicates that a tare weight has been established in the system ZERO Displays when the indicator reading is at Center of Zero lb The displayed weig...

Page 4: ...e provided certain conditions are met2 Print Sends the displayed weight reading to the serial communication port provided certain conditions are met3 1 Scale mustn t be in motion or displaying an erro...

Page 5: ...e Installation Find a suitable location for the indicator and use the included bracket to mount the unit to a wall or table Use this handy guide for mounting the bracket to a wall or table The indicat...

Page 6: ...dicator has been designed to operate from 9 to 12 VDC and ships with an AC wall adapt er designed to operate from the local line voltage All units ship with the appropriate power plug for its area of...

Page 7: ...s Zero Track Band Baud Rate Data Parity Stop Bits Selection 500 to 50000 0 0 5 1 3 5 1200 2400 4800 9600 19200 8_n 7_o 7_E 7_n Level Entering the Setup F Configuration Menu 1 Switch off the indicator...

Page 8: ...0 F3 Zero Track Band Selects the range within which the scale will automatically zero Selections are in display divisions d 0d 0 5d 1d 3d 5d F4 Zero Range Selects the range expressed as a percentage o...

Page 9: ...the procedure Press ZERO key to begin sequence F21 Factory Reset This sub menu will reset all parameters in the F and A menu to the default settings It will not overwrite any previously saved calibra...

Page 10: ...one stop bit 7n 7 data bits with no parity bit and two stop bits 8n 7o 7E 7n A3 Mode of Serial Transmission Selects how data will be sent out of the serial COM port to a printer or computer C Continuo...

Page 11: ...ted in Print Ticket Mode Valid only when A6 is set to 1 Pressing the ZERO key to scroll down one level begins the se quence 0 99 8 A10 Handshaking Enable Enables hardware handshaking for Print Ticket...

Page 12: ...ute the mV V setting for the prompted load weight values Zero Calibration Instructions F16 1 While in the Setup mode scroll to F 16 and then scroll down once using the ZERO down key The display will m...

Page 13: ...alibration test load or check the input data Err2 There is not enough signal from the scale platform to complete the calibration process Most common causes include incorrect weight sensor wiring a mec...

Page 14: ...you to place five identical items onto the scale platform4 3 Place the samples onto the platform all at once and allow the weight indication to stabi lize The screen will change from 5 0 to 5 4 Press...

Page 15: ...ce for remote displays and other remote devices requiring a constant data stream The trans mission automatically occurs at the end of each display update RECOGNIZED HOST COMMANDS ASCII code Hex Symbol...

Page 16: ...xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT Net Start Transmission Carriage Return Line Feed SP SP Space Space pc pieces cu c...

Page 17: ...17 6L 6P 6LB 6PB and CUSTOM Strings A6 3 4 5 6 or 7 These strings are designed for communication with the TSC TDP 225 serial printer...

Page 18: ...tion or load cell dam age due to overloading Err 0 Span Calibration Mode F17 Keyed in weight value is larger than full scale capacity Use a smaller test weight or check keyed in value Err 1 Span Calib...

Page 19: ...d Floods can also damage load cells as well as junction boxes Electromagnetic Interference EMI Call Transcell Tech Support or installer Internal fault Call Transcell Tech Support or installer Transcel...