18
SPECIFICATIONS
Indicator Specifications
Enclosure:
ABS enclosure (IP54)
Display:
0.56" (14 mm) 7-segment, LED, 6 Digit
A-to-D converter
: AD-01
Resolution:
Approximately 200,000 counts @ 3mV/V input
Sampling Rate:
10 Hz
Excitation Voltage
: +5 VDC, 4 x 350
Ω load cells
Input Signal Range
: ±3.125 mV/V
Serial Port:
Full Duplex RS-232C
Operating Temperature
: 14°F to 104°F (-10°C to 40°C)
Power
- 9 Vdc wall adaptor
- DC Power Consumption: 200mA + 15mA/350
Load Cell
- Internal Fuse: 1A 250V SLOW BLOW
ERROR MESSAGES
CODE
MODE
MEANING / POSSIBLE SOLUTION
“oooooo”
Normal Operating
Mode
Gross Overload. A load greater than the rated capacity has been applied
to the scale. Remove the load from the platter. Or try re-calibrating the
scale. Otherwise, check for a bad load cell connection or load cell dam-
age due to overloading.
Err 0
Span Calibration
Mode (F17)
Keyed-in weight value is larger than full-scale capacity. Use a smaller test
weight or check keyed-in value.
Err 1
Span Calibration
Mode (F17)
Keyed-in weight value is less than 1% of full-scale capacity. Use a larger
test weight or check keyed-in value.
Err 2
Span Calibration
Mode (F17)
There is not enough signal from the load cell to produce the internal
counts necessary to properly calibrate the scale. First check all load con-
nections. Use F16 mode to view internal counts.
Err 3
All Modes
Non-volatile memory read error. One or more setup parameters have
been lost.
Err 4
All Modes
Non-volatile memory write error. Indicator needs service.
Err 5
Key-in Span Calibra-
tion Mode (F20)
You have attempted to enter a zero value for C1. Enter a known calibra-
tion value greater than zero.
Err 9
Normal Operating
Mode
Span calibration value has been lost. Re-calibrate the scale.