12
KEY-IN ZERO CALIBRATION VALUE (F19)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss. A
valid zero calibration value, obtained from a successful F16 calibration procedure, must be
used.
1.
While in the Setup mode, scroll to "
F 19
", then scroll down once using the ZERO key. The
display will momentarily show "
ET C 0
", followed by a value of zero
2.
Use the four directional keys to enter in the actual zero calibration value.
3.
After entering the exact value, press the NET/GROSS key to save the value. The display will
show "
E E C 0
" momentarily, and then revert back up to F19.
KEY-IN SPAN CALIBRATION VALUE (F20)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss.
Valid span calibration values, obtained from a successful F17 calibration procedure, must be
used.
1.
While in the Setup mode, scroll to "
F 20
", and then scroll down once using the ZERO key. The
indicator will prompt you to enter the information in Table 3.
2.
If the value shown is correct, press the NET/GROSS key to move to the next parameter.
Otherwise, use the four directional keys to enter in the actual calibration value
3.
After setting the exact value, press the NET/GROSS key to save the value.
4.
If the entered values are entered successfully, the display will show "
E
" momentarily before
continuing to the next parameter. At the completion of the sequence, the indicator will then
revert back up to F20.
CODE
NAME
ET T 1
First Test Weight Value
ET C 1
First Span Calibration Value
ET T 2
Second Test Weight Value
ET C 2
Second Span Calibration Value
ET T 3
Third Test Weight Value
ET C 3
Third Span Calibration Value
Table 3: Calibration Value Entry Table