LEGAL FOR TRADE SEALING
This indicator can be sealed for commercial (Legal for Trade) applications as follows.
1. Power off the indicator.
2. On the back of the indicator, locate the setup/calibration switch cover.
3. Thread a wire security seal through both drilled head screws securing the calibration switch cover
as well as the single drilled head screw holding on the rear panel.